| Literature DB >> 26751290 |
Jochen Klenk1, Gisela Büchele, Ulrich Lindemann, Sabrina Kaufmann, Raphael Peter, Roman Laszlo, Susanne Kobel, Dietrich Rothenbacher.
Abstract
The aim of this study was to assess concurrent validity between activPAL and activPAL3 accelerometers in a sample of 53 community-dwelling older adults ≥ 65 years. Physical activity (PA) was measured simultaneously with activPAL and activPAL3 while performing scripted activities. The level of agreement between both devices was calculated for sitting/lying, standing, and walking. In addition, PA was measured over one week using activPAL to estimate the expected agreement with activPAL3 in real life. Overall agreement between activPAL and activPAL3 was 97%. Compared with activPAL, the largest disagreement was seen for standing, with 5% categorized as walking by activPAL3. For walking and sitting/lying, the disagreement was 2%, respectively. The expected daily differences between activPAL3 and activPAL were +15.0 min (95% CI: 11.3-18.8) for walking and +29.5 min (95% CI: 6.2-52.7) for standing. ActivPAL and activPAL3 showed good agreement in older adults. However, if using these devices interchangeably, observed differences might still bias results.Mesh:
Year: 2016 PMID: 26751290 DOI: 10.1123/japa.2015-0178
Source DB: PubMed Journal: J Aging Phys Act ISSN: 1063-8652 Impact factor: 1.961