| Literature DB >> 26747691 |
Kerstin Zawatzky1, Mirlinda Biba2, Erik L Regalado2, Christopher J Welch3.
Abstract
MISER chromatographic analysis (Multiple Injections in a Single Experimental Run) using supercritical fluid chromatography (SFC) with pressurized carbon dioxide-based eluents is well suited to the high throughput analysis of enantiopurity. SFC is currently the preferred method for fast enantiopurity analysis, with analysis times of only a few seconds achievable in some cases. Injector programming using both the Agilent Infinity and Shimadzu Nexera UC instruments permitted MISER SFC experiments to be performed. Several case studies are presented, showcasing the power and versatility of the technique, with 'plate analysis times' (the time required for analysis of enantiopurity of 96 samples) of less than 33-34 min achievable in the best cases.Entities:
Keywords: Chiral separation; Enantiopurity analysis; High throughput analysis; Multiple Injections in a Single Experimental Run; Supercritical fluid chromatography; Ultrafast enantioseparation
Mesh:
Substances:
Year: 2015 PMID: 26747691 DOI: 10.1016/j.chroma.2015.12.057
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759