Comparison of CAIPIRINHA-VIBE, Radial-VIBE, and conventional VIBE sequences for dynamic contrast-enhanced (DCE) MRI: A validation study using a DCE-MRI phantom.

Bohyun Kim1, Chang Kyung Lee2, Nieun Seo3, Seung Soo Lee3, Jeong Kon Kim3, Yoonseok Choi2, Dong-Cheol Woo2, In Seong Kim4, Dominik Nickel5, Kyung Won Kim6.   

Abstract

OBJECTIVE: To validate radial acquisition of volumetric interpolated breath hold examination (Radial-VIBE) and the controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA-VIBE) sequences for dynamic contrast-enhanced MRI (DCE-MRI) by comparing them to conventional VIBE sequence using a phantom.
METHODS: On a DCE-MRI phantom containing various concentrations of NiCl2 solutions, six minutes of dynamic series and T1 mapping with variable flip angle methods were acquired using conventional VIBE, Radial-VIBE, and CAIPIRINHA-VIBE sequences on 3.0-T scanners. Signal stability and signal linearity were tested for dynamic series and the precision of R1 values were tested for T1 mapping series. The scans were repeatedly performed at two weeks and three months to test repeatability/reproducibility, assessed by within-subject coefficient of variation (WSCV).
RESULTS: Signal stability over six minutes was excellent in all three sequences. Regarding the signal linearity, CAIPIRINHA-VIBE demonstrated the highest linear correlation (r=0.963), followed by conventional VIBE (r=0.959) and Radial-VIBE (r=0.953). Regarding the R1 precision, CAIPIRINHA-VIBE (r=0.985) was the most accurate, followed by conventional VIBE (r=0.861) and Radial-VIBE (r=0.442). CAIPIRINHA-VIBE showed excellent repeatability/reproducibility (WSCV, 1.79-6.71%) compared with Radial-VIBE (WSCV, 2.04-67.2%) and conventional VIBE (WSCV, 3.4-31.9%).
CONCLUSION: In terms of signal stability, signal linearity, R1 precision, and repeatability/reproducibility, CAIPIRINHA-VIBE demonstrated outstanding performance for DCE-MRI compared with Radial-VIBE and conventional VIBE.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dynamic contrast-enhanced; MRI; Parallel imaging; Phantom; Validation

Mesh:

Substances:

Year:  2015        PMID: 26747409     DOI: 10.1016/j.mri.2015.11.011

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  2 in total

Review 1.  Recent advances in parallel imaging for MRI.

Authors:  Jesse Hamilton; Dominique Franson; Nicole Seiberlich
Journal:  Prog Nucl Magn Reson Spectrosc       Date:  2017-05-02       Impact factor: 9.795

2.  The Effects of Breathing Motion on DCE-MRI Images: Phantom Studies Simulating Respiratory Motion to Compare CAIPIRINHA-VIBE, Radial-VIBE, and Conventional VIBE.

Authors:  Chang Kyung Lee; Nieun Seo; Bohyun Kim; Jimi Huh; Jeong Kon Kim; Seung Soo Lee; In Seong Kim; Dominik Nickel; Kyung Won Kim
Journal:  Korean J Radiol       Date:  2017-02-07       Impact factor: 3.500

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.