In vitro aerodynamic characterization of the dose emitted during nebulization of tobramycin high strength solution by novel and jet nebulizer delivery systems.

M Mashat1, B J Clark2, K H Assi2, H Chrystyn3.   

Abstract

BACKGROUND: Chronic infections with Pseudomonas aeruginosa are a leading cause of morbidity in patients with cystic fibrosis (CF). The aim of tobramycin inhalation therapy in CF patients with chronic pulmonary infection is to deliver high amounts of drug directly to the site of infection. TOBI(®) is a tobramycin nebulizer solution (300 mg/5 ml) approved by FDA for maintenance therapy for patient with CF. The 20% tobramycin sulfate solution was reported as the optimal and maximal concentration.
METHODS: Nebulization of high strength tobramycin solution (20% tobramycin sulfate) (HSTS) has been assessed in this study by using different selected high performance nebulizer delivery systems: two different designs of jet nebulizers, and three new nebulizers based on vibrating mesh technology. The aerosol particle size distribution and output characteristics were measured for in vitro performance assessment of the nebulizer systems. The methodology was adapted from the current European standard, EN 13544-1:2001E.
RESULTS: The particle size distribution characteristic measurements showed that all tested nebulizers may be suitable for inhalation of HSTS. The mean (SD) of highest percentage of fine particles (<5 μm) was 77.64 (2.3) % for Sidestream(®), at flow rate 16 L/min. The highest respirable inhaled mass was for Pari LC Plus(®) combined with PariBoyN(®) compressor, with mean (SD) 90.85 (8.6) mg. The mean (SD) of highest drug wastage percentage was 63.9 (3.9) % for Sidestream(®) jet nebulizer combined with compressed air cylinder at flow rate 16 L/min, while the lowest was 2.3 (0.26) % for NE-U22 Omron(®) (high frequency).
CONCLUSIONS: The HSTS can be nebulized by all tested nebulisers but the high frequency NE-U22 Omron(®) and Aeroneb Go(®) are more efficient. When the HSTS compared to TOBI(®), the respirable inhaled dose was increased to more than 73%.
Copyright © 2016. Published by Elsevier Ltd.

Entities:  

Keywords:  CEN method; Cystic fibrosis; High strength tobramycin solution (HSTS); Nebulizer

Mesh:

Substances:

Year:  2015        PMID: 26747025     DOI: 10.1016/j.pupt.2015.12.003

Source DB:  PubMed          Journal:  Pulm Pharmacol Ther        ISSN: 1094-5539            Impact factor:   3.410


  3 in total

1.  Aerosolization Performance of Jet Nebulizers and Biopharmaceutical Aspects.

Authors:  Greta Adorni; Gerrit Seifert; Francesca Buttini; Gaia Colombo; Luciano A Stecanella; Irene Krämer; Alessandra Rossi
Journal:  Pharmaceutics       Date:  2019-08-11       Impact factor: 6.321

2.  Ventilator-associated pneumonia caused by multidrug-resistant Gram-negative bacteria: understanding nebulization of aminoglycosides and colistin.

Authors:  J J Rouby; C Sole-Lleonart; J Rello
Journal:  Intensive Care Med       Date:  2020-01-08       Impact factor: 17.440

3.  Characterisation of Particle Size and Viability of SARS-CoV-2 Aerosols from a Range of Nebuliser Types Using a Novel Sampling Technique.

Authors:  Susan Paton; Simon Clark; Antony Spencer; Isobel Garratt; Ikshitaa Dinesh; Katy-Anne Thompson; Allan Bennett; Thomas Pottage
Journal:  Viruses       Date:  2022-03-19       Impact factor: 5.048

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.