Literature DB >> 26742140

Interference Resilient Sigma Delta-Based Pulse Oximeter.

Mohsen Shokouhian, Richard Morling, Izzet Kale.   

Abstract

Ambient light and optical interference can severely affect the performance of pulse oximeters. The deployment of a robust modulation technique to drive the pulse oximeter LEDs can reduce these unwanted effects and increases the resilient of the pulse oximeter against artificial ambient light. The time division modulation technique used in conventional pulse oximeters can not remove the effect of modulated light coming from surrounding environment and this may cause huge measurement error in pulse oximeter readings. This paper presents a novel cross-coupled sigma delta modulator which ensures that measurement accuracy will be more robust in comparison with conventional fixed-frequency oximeter modulation technique especially in the presence of pulsed artificial ambient light. Moreover, this novel modulator gives an extra control over the pulse oximeter power consumption leading to improved power management.

Mesh:

Year:  2016        PMID: 26742140     DOI: 10.1109/TBCAS.2015.2501359

Source DB:  PubMed          Journal:  IEEE Trans Biomed Circuits Syst        ISSN: 1932-4545            Impact factor:   3.833