A Novel Pairwise Comparison-Based Method to Determine Radiation Dose Reduction Potentials of Iterative Reconstruction Algorithms, Exemplified Through Circle of Willis Computed Tomography Angiography.

Stephan Ellmann1, Ferdinand Kammerer, Michael Brand, Thomas Allmendinger, Matthias S May, Michael Uder, Michael M Lell, Manuel Kramer.   

Abstract

OBJECTIVES: The aim of this study was to determine the dose reduction potential of iterative reconstruction (IR) algorithms in computed tomography angiography (CTA) of the circle of Willis using a novel method of evaluating the quality of radiation dose-reduced images.
MATERIALS AND METHODS: This study relied on ReconCT, a proprietary reconstruction software that allows simulating CT scans acquired with reduced radiation dose based on the raw data of true scans. To evaluate the performance of ReconCT in this regard, a phantom study was performed to compare the image noise of true and simulated scans within simulated vessels of a head phantom. That followed, 10 patients scheduled for CTA of the circle of Willis were scanned according to our institute's standard protocol (100 kV, 145 reference mAs). Subsequently, CTA images of these patients were reconstructed as either a full-dose weighted filtered back projection or with radiation dose reductions down to 10% of the full-dose level and Sinogram-Affirmed Iterative Reconstruction (SAFIRE) with either strength 3 or 5. Images were marked with arrows pointing on vessels of different sizes, and image pairs were presented to observers. Five readers assessed image quality with 2-alternative forced choice comparisons.
RESULTS: In the phantom study, no significant differences were observed between the noise levels of simulated and true scans in filtered back projection, SAFIRE 3, and SAFIRE 5 reconstructions.The dose reduction potential for patient scans showed a strong dependence on IR strength as well as on the size of the vessel of interest. Thus, the potential radiation dose reductions ranged from 84.4% for the evaluation of great vessels reconstructed with SAFIRE 5 to 40.9% for the evaluation of small vessels reconstructed with SAFIRE 3.
CONCLUSIONS: This study provides a novel image quality evaluation method based on 2-alternative forced choice comparisons. In CTA of the circle of Willis, higher IR strengths and greater vessel sizes allowed higher degrees of radiation dose reduction.

Entities:  

Mesh:

Year:  2016        PMID: 26741892     DOI: 10.1097/RLI.0000000000000243

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  7 in total

1.  Dose reduction potential of iterative reconstruction algorithms in neck CTA-a simulation study.

Authors:  Stephan Ellmann; Ferdinand Kammerer; Thomas Allmendinger; Michael Brand; Rolf Janka; Matthias Hammon; Michael M Lell; Michael Uder; Manuel Kramer
Journal:  Dentomaxillofac Radiol       Date:  2016-08-19       Impact factor: 2.419

2.  Quantification of metal artifacts in computed tomography: methodological considerations.

Authors:  Nils Große Hokamp; Brendan Eck; Florian Siedek; Daniel Pinto Dos Santos; Jasmin A Holz; David Maintz; Stefan Haneder
Journal:  Quant Imaging Med Surg       Date:  2020-05

3.  Effect of Radiation Dose Reduction and Reconstruction Algorithm on Image Noise, Contrast, Resolution, and Detectability of Subtle Hypoattenuating Liver Lesions at Multidetector CT: Filtered Back Projection versus a Commercial Model-based Iterative Reconstruction Algorithm.

Authors:  Justin Solomon; Daniele Marin; Kingshuk Roy Choudhury; Bhavik Patel; Ehsan Samei
Journal:  Radiology       Date:  2017-02-07       Impact factor: 11.105

4.  Impact of slice thickness, pixel size, and CT dose on the performance of automatic contouring algorithms.

Authors:  Kai Huang; Dong Joo Rhee; Rachel Ger; Rick Layman; Jinzhong Yang; Carlos E Cardenas; Laurence E Court
Journal:  J Appl Clin Med Phys       Date:  2021-03-29       Impact factor: 2.102

5.  Finding the optimal tube current and iterative reconstruction strength in liver imaging; two needles in one haystack.

Authors:  Bibi Martens; Joris G A Bosschee; Sander M J Van Kuijk; Cécile R L P N Jeukens; Maikel T H Brauer; Joachim E Wildberger; Casper Mihl
Journal:  PLoS One       Date:  2022-04-07       Impact factor: 3.240

6.  Low-Dose Abdominal CT Using a Deep Learning-Based Denoising Algorithm: A Comparison with CT Reconstructed with Filtered Back Projection or Iterative Reconstruction Algorithm.

Authors:  Yoon Joo Shin; Won Chang; Jong Chul Ye; Eunhee Kang; Dong Yul Oh; Yoon Jin Lee; Ji Hoon Park; Young Hoon Kim
Journal:  Korean J Radiol       Date:  2020-03       Impact factor: 3.500

7.  Are quantitative features of lung nodules reproducible at different CT acquisition and reconstruction parameters?

Authors:  Barbaros S Erdal; Mutlu Demirer; Kevin J Little; Chiemezie C Amadi; Gehan F M Ibrahim; Thomas P O'Donnell; Rainer Grimmer; Vikash Gupta; Luciano M Prevedello; Richard D White
Journal:  PLoS One       Date:  2020-10-15       Impact factor: 3.240

  7 in total