Failure rate and conductor externalization in the Biotronik Linox/Sorin Vigila implantable cardioverter-defibrillator lead.

Fabian Noti1, Anna Lam1, Nicole Klossner1, Jens Seiler1, Helge Servatius1, Argelia Medeiros-Domingo1, Vân Nam Tran1, Andreas Haeberlin1, Juerg Fuhrer1, Hildegard Tanner1, Laurent Roten2.   

Abstract

BACKGROUND: We observed a case of conductor externalization in a Biotronik Linox lead.
OBJECTIVE: The purpose of this study was to investigate lead performance of the Linox lead and the identical Sorin Vigila lead and prevalence of conductor externalization.
METHODS: We compared lead performance of all Linox and Vigila leads implanted at our center (BL group; n = 93) with that of all Boston Scientific Endotak Reliance leads (ER group; n = 190) and Medtronic Sprint Quattro leads (SQ group; n = 202) implanted during the same period. We screened all patients in the BL group for conductor externalization.
RESULTS: We identified 8 cases of lead failures in the BL group (index case of conductor externalization, 6 cases of nonphysiological high-rate sensing, and 1 case of high-voltage conductor fracture). Prospective fluoroscopic screening of 98% of all active BL group cases revealed 1 additional case of conductor externalization. The median follow-up was 41, 27, and 29 months for the BL group, ER group, and SQ group, respectively; lead survival was 94.9%, 99.2%, and 100% at 3 years and 88%, 97.5%, and 100% at 5 years (P = .038 for BL group vs ER group and P = .007 for BL group vs SQ group using the log-rank test). Younger age at implant was an independent predictor of lead failure in the BL group (adjusted hazard ratio 0.85; 95% confidence interval 0.77-0.94; P = .001).
CONCLUSION: At our center, survival of the Linox lead is 88% at 5 years and significantly worse than that of other leads. Conductor externalization is present in a minority of failed Linox leads. Younger age at implant is an independent predictor of Linox lead failure.
Copyright © 2016 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Conductor externalization; Defibrillator lead; Endotak Reliance; Lead failure; Linox; Sprint Quattro

Mesh:

Year:  2015        PMID: 26738946     DOI: 10.1016/j.hrthm.2015.12.038

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  3 in total

1.  Very-late onset twiddler syndrome as an unusual cause of syncope.

Authors:  Andreas Haeberlin; Rémi Chauvel; Antoine Noel; Nicolas Welte; Philippe Ritter; Pierre Bordachar; Sylvain Ploux
Journal:  J Interv Card Electrophysiol       Date:  2019-08-24       Impact factor: 1.900

2.  First time description of early lead failure of the Linox Smart lead compared to other contemporary high-voltage leads.

Authors:  Vanessa Weberndörfer; Tobias Nyffenegger; Ian Russi; Miriam Brinkert; Benjamin Berte; Stefan Toggweiler; Richard Kobza
Journal:  J Interv Card Electrophysiol       Date:  2018-05-01       Impact factor: 1.900

Review 3.  Externalized conductors and insulation failure in Biotronik defibrillator leads: History repeating or a false alarm?

Authors:  Elia De Maria; Ambra Borghi; Lorenzo Bonetti; Pier Luigi Fontana; Stefano Cappelli
Journal:  World J Clin Cases       Date:  2017-02-16       Impact factor: 1.337

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.