Design optimization of neuroendoscopic continuum instruments for third ventriculostomy and tumor biopsy.

Kyle W Eastwood, Thomas Looi, Hani E Naguib, James M Drake.   

Abstract

A simulation-based approach to the design of procedure-specific dexterous neuroendoscopic continuum instruments for endoscopic third ventriculostomy and tumor biopsy is proposed. Given pre-operative CT and MRI images, the algorithm returns instrument design specifications including lengths, curvatures and alternative positions for the surgical incision while respecting anatomical boundaries. This methodology proposes a novel clinically-guided geometric representation of surgical targets specific to this indication. The additional constraints imposed by the presence of surgical trocars and endoscopes are also considered. Finally, we conclude with a clinical example to demonstrate the proposed approach.

Entities:  

Mesh:

Year:  2015        PMID: 26737380     DOI: 10.1109/EMBC.2015.7319480

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  2 in total

1.  A multiport MR-compatible neuroendoscope: spanning the gap between rigid and flexible scopes.

Authors:  Sunil Manjila; Margherita Mencattelli; Benoit Rosa; Karl Price; Georgios Fagogenis; Pierre E Dupont
Journal:  Neurosurg Focus       Date:  2016-09       Impact factor: 4.047

2.  Towards FBG-Based Shape Sensing for Micro-scale and Meso-Scale Continuum Robots with Large Deflection.

Authors:  Yash Chitalia; Nancy Joanna Deaton; Seokhwan Jeong; Nahian Rahman; Jaydev P Desai
Journal:  IEEE Robot Autom Lett       Date:  2020-01-28
  2 in total