Power line interference attenuation in multi-channel sEMG signals: Algorithms and analysis.

S D H Soedirdjo, K Ullah, R Merletti.   

Abstract

Electromyogram (EMG) recordings are often corrupted by power line interference (PLI) even though the skin is prepared and well-designed instruments are used. This study focuses on the analysis of some of the recent and classical existing digital signal processing approaches have been used to attenuate, if not eliminate, the power line interference from EMG signals. A comparison of the signal to interference ratio (SIR) of the output signals is presented, for four methods: classical notch filter, spectral interpolation, adaptive noise canceller with phase locked loop (ANC-PLL) and adaptive filter, applied to simulated multichannel monopolar EMG signals with different SIR. The effect of each method on the shape of the EMG signals is also analyzed. The results show that ANC-PLL method gives the best output SIR and lowest shape distortion compared to the other methods. Classical notch filtering is the simplest method but some information might be lost as it removes both the interference and the EMG signals. Thus, it is obvious that notch filter has the lowest performance and it introduces distortion into the resulting signals.

Mesh:

Year:  2015        PMID: 26737127     DOI: 10.1109/EMBC.2015.7319227

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  3 in total

1.  Characterization of Forearm Muscle Activation in Duchenne Muscular Dystrophy via High-Density Electromyography: A Case Study on the Implications for Myoelectric Control.

Authors:  Kostas Nizamis; Noortje H M Rijken; Robbert van Middelaar; João Neto; Bart F J M Koopman; Massimo Sartori
Journal:  Front Neurol       Date:  2020-04-15       Impact factor: 4.003

2.  An Ultra-Low Power Surface EMG Sensor for Wearable Biometric and Medical Applications.

Authors:  Yi-Da Wu; Shanq-Jang Ruan; Yu-Hao Lee
Journal:  Biosensors (Basel)       Date:  2021-10-21

3.  Virtual Sensor of Surface Electromyography in a New Extensive Fault-Tolerant Classification System.

Authors:  Karina de O A de Moura; Alexandre Balbinot
Journal:  Sensors (Basel)       Date:  2018-05-01       Impact factor: 3.576

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.