Literature DB >> 26736606

Wearable lung-health monitoring system with electrical impedance tomography.

Sunjoo Hong, Jaehyuk Lee, Hoi-Jun Yoo.   

Abstract

The wearable lung-health monitoring system is proposed with an electrical impedance tomography (EIT). The proposed system has light belt-type form factor which is implemented with the EIT integrated circuit (IC) on the planar-fashionable circuit board (P-FCB) technology. The EIT IC provides programmable current stimulation which is optimally controlled by the results of contact impedance monitoring. The measured data is transmitted to the mobile device and the lung EIT images are reconstructed and displayed with up to 20 frames/s real-time. From the lung EIT image, the measured lung air volume ratio can be used as an indicator of the lung-health, and other various parameters can be extracted to monitor lung status. The proposed wearable system achieves the user convenience for lung-health monitoring which can be used personally at home. The proposed system is fully implemented and verified on both in-vitro and in-vivo tests.

Mesh:

Year:  2015        PMID: 26736606     DOI: 10.1109/EMBC.2015.7318706

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Simple Wireless Impedance Pneumography System for Unobtrusive Sensing of Respiration.

Authors:  Pablo Aqueveque; Britam Gómez; Emyrna Monsalve; Enrique Germany; Paulina Ortega-Bastidas; Sebastián Dubo; Esteban J Pino
Journal:  Sensors (Basel)       Date:  2020-09-14       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.