Literature DB >> 26732420

One-Year Outcomes in "Real-World" Patients Treated With a Thin-Strut, Platinum-Chromium, Everolimus-Eluting Stent (from the PROMUS Element Plus US Post-Approval Study [PE-Plus PAS]).

David E Kandzari1, Nima Amjadi2, Christopher Caputo3, Steven K Rowe4, Jerome Williams5, Hoshedar P Tamboli6, Thomas Christen7, Dominic J Allocco7, Keith D Dawkins7.   

Abstract

The PROMUS Element Plus US Post-Approval Study (PE-Plus PAS) was a prospective, open-label, multicenter, observational study designed to examine outcomes in everyday clinical practice in patients treated with everolimus-eluting, platinum-chromium PROMUS Element Plus stents at 52 centers in the United States. This is the first report of results from this large study. The primary end point of the PE-Plus PAS was 12-month cardiac death or myocardial infarction in the more restricted population of "PLATINUM-like" patients pooled from the PE-Plus PAS, PE-PROVE (PROMUS Element European post-approval study), and PLATINUM Workhorse/Small Vessel trials. Additional clinical end points were tested in the overall PE-Plus PAS patient population. Of the 2,683 patients enrolled in PE-Plus PAS, 70% were men, mean age was 64 years, 33% had diabetes, and 29% were "PLATINUM-like." Among the PLATINUM-like patients, 12-month cardiac death or myocardial infarction was 1.8% (33 of 1,855) with an upper 1-sided 95% confidence interval of 2.3%, which was significantly less than the prespecified performance goal of 3.2% (pnoninferiority <0.001). In the overall PE-Plus population, 12-month target vessel failure (defined as death, MI, or revascularization related to the target vessel) was 6.7% (170 of 2,554), cardiac death was 1.4% (37 of 2,554), MI was 1.1% (28 of 2,554), and ARC-definite/probable stent thrombosis was 0.7% (19 of 2,554). A prespecified secondary end point of 12-month target vessel failure in diabetic patients demonstrated a rate of 4.2% (14 of 332) with an upper 1-sided 95% confidence interval of 6.03%, which was significantly less than the performance goal of 12.6% (pnoninferiority <0.001). In conclusion, in this large registry of unselected patients, coronary artery revascularization with the PROMUS Element Plus everolimus-eluting stent demonstrates favorable results with low 1-year clinical event rates.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26732420     DOI: 10.1016/j.amjcard.2015.11.043

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  4 in total

1.  Outcomes in Women and Minorities Compared With White Men 1 Year After Everolimus-Eluting Stent Implantation: Insights and Results From the PLATINUM Diversity and PROMUS Element Plus Post-Approval Study Pooled Analysis.

Authors:  Wayne Batchelor; David E Kandzari; Scott Davis; Luis Tami; John C Wang; Islam Othman; Osvaldo S Gigliotti; Amir Haghighat; Sarabjeet Singh; Mario Lopez; Gregory Giugliano; Phillip A Horwitz; Jaya Chandrasekhar; Paul Underwood; Craig A Thompson; Roxana Mehran
Journal:  JAMA Cardiol       Date:  2017-12-01       Impact factor: 14.676

2.  Does study subject diversity influence cardiology research site performance?: Insights from 2 U.S. National Coronary Stent Registries.

Authors:  Wayne B Batchelor; Abdulla A Damluji; Celina Yong; Mona Fiuzat; Scott D Barnett; David E Kandzari; Matthew W Sherwood; Kelly C Epps; Behnam N Tehrani; Dominic J Allocco; Ian T Meredith; JoAnn Lindenfeld; Christopher M O'Connor; Roxana Mehran
Journal:  Am Heart J       Date:  2021-02-24       Impact factor: 5.099

Review 3.  Sex-Specific Outcomes in Cardiovascular Device Evaluations.

Authors:  Mohammed Imran Ghare; Daniela Tirziu; Jinnette Dawn Abbott; Elissa Altin; Yiping Yang; Vivian Ng; Cindy Grines; Alexandra Lansky
Journal:  J Womens Health (Larchmt)       Date:  2020-06-15       Impact factor: 3.017

4.  Safety and efficacy of Everolimus-Eluting bioabsorbable Polymer-Coated stent in patients with long coronary lesions: The EVOLVE 48 study.

Authors:  Dimitrios Karmpaliotis; Robert Stoler; Simon Walsh; Seif El-Jack; Srinivasa Potluri; Jeffrey Moses; Keith Oldroyd; Adrian Banning; Mark Webster; Azfar Zaman; Willis Wu; Mudassar Ahmed; Paul Underwood; Dominic Allocco
Journal:  Catheter Cardiovasc Interv       Date:  2021-05-29       Impact factor: 2.585