| Literature DB >> 26726935 |
J H Sun1, F Y Guan2, X F Zhu3, Z W Ning3, T J Ma4, J H Liu3, T Deng4.
Abstract
In this work, a micro packed gas chromatograph column integrated with a micro heater was fabricated by using laser etching technology (LET) for analyzing environmental gases. LET is a powerful tool to etch deep well-shaped channels on the glass wafer, and it is the most effective way to increase depth of channels. The fabricated packed GC column with a length of over 1.6m, to our best knowledge, which is the longest so far. In addition, the fabricated column with a rectangular cross section of 1.2mm (depth) × 0.6mm (width) has a large aspect ratio of 2:1. The results show that the fabricated packed column had a large sample capacity, achieved a separation efficiency of about 5800 plates/m and eluted highly symmetrical Gaussian peaks.Keywords: Laser etching technology; Micro gas chromatography system; Micro packed gas chromatography column; Separation efficiency
Mesh:
Year: 2015 PMID: 26726935 DOI: 10.1016/j.chroma.2015.12.001
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759