| Literature DB >> 26724073 |
Robert Gill1, Evan Bush1, Philipp Haueter2, Peter Loutzenhiser1.
Abstract
A systematic methodology for characterizing a novel and newly fabricated high-flux solar simulator is presented. The high-flux solar simulator consists of seven xenon short-arc lamps mounted in truncated ellipsoidal reflectors. Characterization of spatial radiative heat flux distribution was performed using calorimetric measurements of heat flow coupled with CCD camera imaging of a Lambertian target mounted in the focal plane. The calorimetric measurements and images of the Lambertian target were obtained in two separate runs under identical conditions. Detailed modeling in the high-flux solar simulator was accomplished using Monte Carlo ray tracing to capture radiative heat transport. A least-squares regression model was used on the Monte Carlo radiative heat transfer analysis with the experimental data to account for manufacturing defects. The Monte Carlo ray tracing was calibrated by regressing modeled radiative heat flux as a function of specular error and electric power to radiation conversion onto measured radiative heat flux from experimental results. Specular error and electric power to radiation conversion efficiency were 5.92 ± 0.05 mrad and 0.537 ± 0.004, respectively. An average radiative heat flux with 95% errors bounds of 4880 ± 223 kW ⋅ m(-2) was measured over a 40 mm diameter with a cavity-type calorimeter with an apparent absorptivity of 0.994. The Monte Carlo ray-tracing resulted in an average radiative heat flux of 893.3 kW ⋅ m(-2) for a single lamp, comparable to the measured radiative heat fluxes with 95% error bounds of 892.5 ± 105.3 kW ⋅ m(-2) from calorimetry.Entities:
Year: 2015 PMID: 26724073 DOI: 10.1063/1.4936976
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523