Validity of sidestream endtidal carbon dioxide measurement in critically ill, mechanically ventilated children.

Hylke H A C M van der Heijden1, Gerben J Truin2, Joyce Verhaeg1, Peggy van der Pol3, Joris Lemson1.   

Abstract

INTRODUCTION: Capnography is used to monitor the endtidal carbon dioxide tension (EtCO2) in exhaled gas. Sidestream capnography has great potential to monitor mechanically ventilated pediatric patients, given the continuous sampling from the endotracheal tube into a gas sensor. However, hemodynamic and respiratory impairments may reduce reliability and validity of sidestream capnography to monitor arterial carbon dioxide tension (PaCO2) in critically ill, mechanically ventilated children.
METHODS: In 47 mechanically ventilated pediatric patients (aged 0-14 years, median age 17.2 months), a total of 341 consecutive measurements of PaCO2, EtCO2, respiratory and hemodynamic parameters were performed, and capnogram shape was determined. Validity was assessed with the Bland-Altman limit of agreement (loa), mixed models were used to adjust for variation between patients, and potential confounders were considered with multivariable analyses.
RESULTS: EtCO2 (mean 4.50 ± 0.96 kPa) underestimated PaCO2 (mean 5.37 ± 0.87) considerably, resulting in a loa of 0.87 kPa [95% confidence interval (95% CI) -1.03;2.77] and 42.2% percentage error. The association improved significantly b = 0.54 [95 %CI = 0.45;0.64, P < 0.001] when corrected for individual differences. The association between EtCO2 and PaCO2 was not influenced by any of the potential confounders.
CONCLUSIONS: Sidestream capnography in mechanically ventilated infants and children seems moderately reliable and valid when corrected for individual differences. Therefore, it could only be used with caution for trend estimation in the individual patient.
© 2015 John Wiley & Sons Ltd.

Entities:  

Keywords:  capnography; children; reliability; validity

Mesh:

Substances:

Year:  2015        PMID: 26714621     DOI: 10.1111/pan.12827

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  1 in total

1.  Comparison of Simultaneous Capillary Blood Gases and End-Tidal Carbon Dioxide in Mechanically Ventilated Pediatric Patients with Acute Respiratory Failure.

Authors:  Capan Konca; Mehmet Tekin; Fatih Uckardes; Samet Benli; Ahmet Kucuk
Journal:  J Pediatr Intensive Care       Date:  2020-07-20
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.