Literature DB >> 26711310

Array-in-well platform-based multiplex assay for the simultaneous detection of anti-HIV- and treponemal-antibodies, and Hepatitis B surface antigen.

Sheikh M Talha1, Petri Saviranta2, Liisa Hattara2, Tytti Vuorinen3, Jukka Hytönen4, Navin Khanna5, Kim Pettersson6.   

Abstract

Multiplex assays detecting sets of related clinical analytes simultaneously can save considerable amount of time and resources. Array-in-well (AIW) is a powerful platform for the multiplex detection of different analytes where microarrays can be printed at the bottom of microtiter wells, thus combining the potential of microarrays with the ease of handling microtiter wells. We have developed a single-step AIW assay for the simultaneous screening of HIV, Treponema pallidum subspecies pallidum (causing syphilis) and Hepatitis B virus infections targeting the specific detection of anti-HIV- and treponemal-antibodies and Hepatitis B surface antigen (HBsAg), respectively, using two different fluorescent label technologies i.e. DyLight 633 and europium nanoparticle. Double-antigen assay formats were used for anti-HIV- and treponemal-antibody detection that can simultaneously detect both IgG and IgM, and thus reduce the window period of detection. AIW assay was evaluated with well characterized serum/plasma samples (n=111), and the qualitative results were in near complete agreement with those of the reference assays. The AIW assay exhibited 100% sensitivities for all three analytes, and 100% specificities for anti-HIV antibodies and HBsAg, and 98.6% specificity for treponemal antibodies. The limit of detection of HBsAg in AIW assay was 0.18 ng/ml. This high performing AIW assay has the potential to be used as a multiplex screening test for these three infections.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Array-in-well; Diagnostics; HIV; Hepatitis B virus; Multiplex; Syphilis

Mesh:

Substances:

Year:  2015        PMID: 26711310     DOI: 10.1016/j.jim.2015.12.007

Source DB:  PubMed          Journal:  J Immunol Methods        ISSN: 0022-1759            Impact factor:   2.303


  2 in total

1.  Double-Antigen Lateral Flow Immunoassay for the Detection of Anti-HIV-1 and -2 Antibodies Using Upconverting Nanoparticle Reporters.

Authors:  Iida Martiskainen; Etvi Juntunen; Teppo Salminen; Karoliina Vuorenpää; Sherif Bayoumy; Tytti Vuorinen; Navin Khanna; Kim Pettersson; Gaurav Batra; Sheikh M Talha
Journal:  Sensors (Basel)       Date:  2021-01-06       Impact factor: 3.576

2.  Towards the Development of a 3-D Biochip for the Detection of Hepatitis C Virus.

Authors:  Mariia Antipchik; Dmitry Polyakov; Ekaterina Sinitsyna; Apollinariia Dzhuzha; Mikhail Shavlovsky; Evgenia Korzhikova-Vlakh; Tatiana Tennikova
Journal:  Sensors (Basel)       Date:  2020-05-10       Impact factor: 3.576