Literature DB >> 26710045

Stryker SMARTLock Hybrid Maxillomandibular Fixation System: Clinical Application, Complications, and Radiographic Findings.

Douglas E Kendrick1, Chan M Park, Jesse M Fa, Jacob S Barber, A Thomas Indresano.   

Abstract

BACKGROUND: The SMARTLock Hybrid MMF System from Stryker is a newer approach for maxillomandibular fixation. This study was performed to determine the clinical application, complications, radiographic findings, and cost effectiveness of the SMARTLock system.
METHODS: A retrospective cohort study was performed with the SMARTLock system over 6 months. Demographics, history, fracture location, placement/removal time, and complications were obtained, along with cost analysis.
RESULTS: The authors identified 35 patients with the SMARTLock system. Twenty-four patients remained after exclusion criteria. There were 19 male patients (79 percent) and five female patients (21 percent), with a mean age of 30.7 years. The mean application time of the SMARTLock system was 14.4 minutes, and the mean removal time was 10.5 minutes. Three hundred nineteen total screws were placed. The number and percentage of patients with complications associated with the SMARTLock system were as follows: mucosal overgrowth [n = 9 (38 percent)], screw loosening [n = 4 (17 percent)], lip irritation [n = 4 (17 percent)], malocclusion [n = 3 (13 percent)], nonunion [n = 1 (4 percent)], wound dehiscence [n = 1 (4 percent)], screw loss [n = 1 (4 percent)], tooth devitalization [n = 1 (4 percent)], loose plate [n = 1 (4 percent)], and plate fracture [n = 1 (4 percent)]. There were no instances of sharps exposure, tooth loss, or infection. One tooth required endodontic therapy. The number of screws that damaged teeth on cone-beam computed tomographic imaging was 24 (7.5 percent). The cost analysis showed similar cost between Erich arch bars and the SMARTLock system.
CONCLUSIONS: This study suggests that the SMARTLock Hybrid MMF System is safe and easy to use, and with a cost similar to that of Erich arch bars. Appropriate treatment planning and previous surgical experience should be used to determine appropriate case selection, as this system is not ideal in all situations. CLINICAL QUESTION/LEVEL OF EVIDENCE: Therapeutic, IV.

Entities:  

Mesh:

Year:  2016        PMID: 26710045     DOI: 10.1097/PRS.0000000000001920

Source DB:  PubMed          Journal:  Plast Reconstr Surg        ISSN: 0032-1052            Impact factor:   4.730


  6 in total

1.  A Clinical Comparison and Economic Evaluation of Erich Arch Bars, 4-Point Fixation, and Bone-Supported Arch Bars for Maxillomandibular Fixation.

Authors:  Meade C Edmunds; T Alex McKnight; Christopher M Runyan; Brian W Downs; Jordan L Wallin
Journal:  JAMA Otolaryngol Head Neck Surg       Date:  2019-06-01       Impact factor: 6.223

2.  The Hybrid Arch Bar Is a Cost-Beneficial Alternative in the Open Treatment of Mandibular Fractures.

Authors:  Renata Khelemsky; David Powers; Seth Greenberg; Visakha Suresh; Ellen Johnson Silver; Michael Turner
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2018-04-12

3.  The use of Synthes MatrixWAVE bone anchored arch bars for closed treatment of multiple concurrent mandibular fractures.

Authors:  Elizabeth Kiwanuka; Rajiv Iyengar; Charles Christopher Jehle; Raman Mehrzad; Daniel Kwan
Journal:  J Oral Biol Craniofac Res       Date:  2017-09-01

4.  A Technique for Reduction of Edentulous Fractures Using Dentures and SMARTLock Hybrid Fixation System.

Authors:  Anna Rose Carlson; Ronnie Labib Shammas; Alexander Christopher Allori; David Bryan Powers
Journal:  Plast Reconstr Surg Glob Open       Date:  2017-09-05

5.  The Use of Screw Retained Hybrid Arch Bar for Maxillomandibular Fixation in the Treatment of Mandibular Fractures: A Comparative Study.

Authors:  Saif T Hamid; Salwan Y Bede
Journal:  Ann Maxillofac Surg       Date:  2022-02-01

6.  Dental occlusion ties: A rapid, safe, and non-invasive maxillo-mandibular fixation technology.

Authors:  Alan W Johnson
Journal:  Laryngoscope Investig Otolaryngol       Date:  2017-04-12
  6 in total