Susanna Ceffa1, Richard Luhanga2, Mauro Andreotti3, Davide Brambilla1, Fulvio Erba4, Haswel Jere2, Sandro Mancinelli5, Marina Giuliano6, Leonardo Palombi5, Maria Cristina Marazzi7.
Abstract
Assessing treatment efficacy and early infant diagnosis (EID) are critical issues in HIV disease management. Point-of-care assays may greatly increase the possibility to access laboratory monitoring also in rural areas. Recently two new laboratory tests have been developed by Cepheid (Sunnyvale, California) the Xpert HIV-1 Viral Load for viral load determination and the Xpert HIV-1 Qualitative for early infant diagnosis. We conducted a study in Blantyre, Malawi, comparing the 2 methods versus the Abbott real time quantitative and qualitative assays, for viral load and EID respectively. We tested 300 plasma samples for viral load determination and 200 samples for infant diagnosis. HIV-1 RNA values of the 274 samples quantified by both assays were highly correlated (Pearson r=0.95, R(2)=0.90). In 90.9% of the cases the two methods were concordant in defining the HIV-1 RNA levels as detectable or undetectable. For EID, the Xpert HIV-1 Qualitative assay yielded the same identical results as the Abbott assay. Both the quantitative and the qualitative Xpert assays are promising tools to monitor treatment efficacy in HIV patients receiving treatment and for early diagnosis in HIV-exposed infants.Entities:
Keywords: Early infant diagnosis; HIV-1; POC tests; Viral load
Mesh:
Substances:
Year: 2015 PMID: 26709099 DOI: 10.1016/j.jviromet.2015.12.007
Source DB: PubMed Journal: J Virol Methods ISSN: 0166-0934 Impact factor: 2.014