Effective dose reduction in spine radiographic imaging by choosing the less radiation-sensitive side of the body.

Avi Ben-Shlomo1, Gabriel Bartal2, Morris Mosseri3, Boaz Avraham2, Yosef Leitner4, Shay Shabat4.   

Abstract

BACKGROUND CONTEXT: X-ray absorption is highest in the organs and tissues located closest to the radiation source. The photon flux that crosses the body decreases from the entry surface toward the image receptor. The internal organs absorb x-rays and shield each other during irradiation. Therefore, changing the x-ray projection angle relative to the patient for specific spine procedures changes the radiation dose that each organ receives. Every organ has different radiation sensitivity, so irradiation from different sides of the body changes the biological influence and radiation risk potential on the total body, that is the effective dose (ED).
PURPOSE: The study aimed to determine the less radiation-sensitive sides of the body during lateral and anterior-posterior (AP) or posterior anterior (PA) directions. STUDY
DESIGN: The study used exposure of patient phantoms and Monte Carlo simulation of the effective doses. PATIENT SAMPLE: Calculations for adults and 10-year-old children were included because the pediatric population has a greater lifetime radiation risk than adults. OUTCOME MEASURES: Pediatric and adult tissue and organ doses and ED from cervical, thoracic, and lumbar x-ray spine examinations were performed from different projections.
METHODS: Standard mathematical phantoms for adults and 10-year-old children, using PCXMC 2.0 software based on Monte Carlo simulations, were used to calculate pediatric and adult tissue and organ doses and ED. The study was not funded. The authors have no conflicts of interest to declare.
RESULTS: Spine x-ray exposure from various right (RT) LAT projection angles was associated with lower ED compared with the same left (LT) LAT projections (up to 28% and 27% less for children aged 10 and adults, respectively). The PA spine projections showed up to 64% lower ED for children aged 10 and 65% for adults than AP projections. The AP projection at the thoracic spine causes an excess breast dose of 543.3% and 597.0% for children aged 10 and adults, respectively.
CONCLUSIONS: Radiation ED in spine procedures can be significantly reduced by performing x-ray exposures through the less radiation-sensitive sides of the body, which are PA in the frontal position and right lateral in the lateral position.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Dosimetry; Effective dose; Positioning; Projection; Radiography; Spine

Mesh:

Year:  2015        PMID: 26704861     DOI: 10.1016/j.spinee.2015.12.012

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  5 in total

1.  Head and neck effective dose and quantitative assessment of image quality: a study to compare cone beam CT and multislice spiral CT.

Authors:  Cosimo Nardi; Cinzia Talamonti; Stefania Pallotta; Paola Saletti; Linda Calistri; Cesare Cordopatri; Stefano Colagrande
Journal:  Dentomaxillofac Radiol       Date:  2017-06-20       Impact factor: 2.419

2.  Lumbar spine radiographs - is it time for widespread adoption of posteroanterior projection?

Authors:  Christopher Green; Guru Karnati; Katharine Thomson; Ashok Subramanian
Journal:  Br J Radiol       Date:  2019-08-07       Impact factor: 3.039

3.  Comparing the supine and erect pelvis radiographic examinations: an evaluation of anatomy, image quality and radiation dose.

Authors:  Kevin Flintham; Kholoud Alzyoud; Andrew England; Peter Hogg; Beverly Snaith
Journal:  Br J Radiol       Date:  2021-05-14       Impact factor: 3.629

4.  Comparison of anteroposterior and posteroanterior projection in lumbar spine radiography.

Authors:  Erna Alukic; Damijan Skrk; Nejc Mekis
Journal:  Radiol Oncol       Date:  2018-05-14       Impact factor: 2.991

5.  Effective dose and image optimisation of lateral lumbar spine radiography: a phantom study.

Authors:  Zer Hau Lai; Cláudia Sá Dos Reis; Zhonghua Sun
Journal:  Eur Radiol Exp       Date:  2020-02-13
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.