Validation of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors in a general adult population according to the European Society of Hypertension and the ANSI/AAMI/ISO 81060-2: 2013 protocols.

Sen Bing1, Kang Chen, Hong Hou, Weijuan Zhang, Linyi Li, Jiao Wei, Chang Shu, Yi Wan.   

Abstract

OBJECTIVE: This study aimed to determine the accuracy of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors according to the European Society of Hypertension International Protocol revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols.
METHODS: The devices were assessed on 33 participants according to the European Society of Hypertension requirements and were then tested on 85 participants according to the ANSI/AAMI/ISO 81060-2:2013 criteria. Procedures and data analysis were carried out following protocol guidelines precisely.
RESULTS: The Microlife BP A200 Comfort and W2 Slim devices passed the criteria of the European Society of Hypertension International Protocol revision 2010 for both systolic blood pressure and diastolic blood pressure. The devices also fulfilled the ANSI/AAMI/ISO 81060-2:2013 criteria, with mean differences of SBP and DPB between the devices and observers of 0.38±5.12 and 0.28±4.29 mmHg for the BP A200 Comfort and 1.01±6.80 and 0.34±5.62 mmHg for the W2 Slim, respectively.
CONCLUSION: The Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors fulfilled the European Society of Hypertension revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols, and can be recommended for self-measurement in the general population.

Entities:  

Mesh:

Year:  2016        PMID: 26683381     DOI: 10.1097/MBP.0000000000000169

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  4 in total

1.  Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS).

Authors:  Farahnaz Joukar; Sara Yeganeh; Mohammadreza Naghipour; Soheil Hassanipour; Hossein-Ali Nikbakht; Fariborz Mansour-Ghanaei
Journal:  Med Devices (Auckl)       Date:  2020-09-03

2.  The QardioArm App in the Assessment of Blood Pressure and Heart Rate: Reliability and Validity Study.

Authors:  Victoria Mazoteras Pardo; Marta E Losa Iglesias; José López Chicharro; Ricardo Becerro de Bengoa Vallejo
Journal:  JMIR Mhealth Uhealth       Date:  2017-12-15       Impact factor: 4.773

3.  The QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010.

Authors:  Victoria Mazoteras-Pardo; Ricardo Becerro-De-Bengoa-Vallejo; Marta Elena Losa-Iglesias; Daniel López-López; Patricia Palomo-López; David Rodríguez-Sanz; César Calvo-Lobo
Journal:  JMIR Mhealth Uhealth       Date:  2018-10-25       Impact factor: 4.773

4.  Associations between cardiorespiratory fitness, fatness, hemodynamic characteristics, and sedentary behaviour in primary school-aged children.

Authors:  Garyfallia Pepera; Savvas Hadjiandrea; Ilias Iliadis; Gavin R H Sandercock; Ladislav Batalik
Journal:  BMC Sports Sci Med Rehabil       Date:  2022-02-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.