Literature DB >> 26682769

Dosimetric Performance of Newly Developed Farmer-Type Ionization Chamber in Radiotherapy Practice.

Sathiyan Saminathan1, Henry Finlay Godson2, Retna Ponmalar2, Ravikumar Manickam2, James Mazarello3, Rahul Fernandes3.   

Abstract

Dose measurement with ionization chamber is essential to deliver accurate dose to the tumor in radiotherapy. The cylindrical Farmer-type ionization chamber is recommended by various dosimetry protocols for dose measurement of radiotherapy beams. The air-equivalent graphite wall Farmer-type ionization chamber (FAR 65 GB) of active volume 0.65 cm3 with aluminum as the central electrode material was fabricated. Various dosimetric parameters were studied for the newly developed ionization chamber in cobalt-60, 6 and 18 MV photon beams. The preirradiation and postirradiation leakage of the chamber was within 0.08%. The long-term stability and the stem effect of the chamber were within 0.07% and 0.3%, respectively. The sensitivity of the ionization chamber was found to be 22.15 nC/Gy. The chamber shows linear response with dose for cobalt-60, 6 and 18 MV photon beams. The ion recombination correction factor increases with increase in bias voltage. For all energies and field sizes, the polarity correction factor is almost closer to unity. The ion recombination and polarity correction measurements show that the polarizing potential and polarity recommended during the calibration of ionization chamber should be used for routine measurement to avoid the uncertainty. The chamber response is independent of dose rate and energy. The chamber is cost-effective and shows precise and reproducible response. The study carried out confirms that the newly fabricated ion chamber can be used in the measurement of absolute dose for high-energy photon beams.
© The Author(s) 2015.

Entities:  

Keywords:  absolute dosimetry; chamber characteristics; farmer-type chamber; photon beams

Mesh:

Substances:

Year:  2015        PMID: 26682769     DOI: 10.1177/1533034615621635

Source DB:  PubMed          Journal:  Technol Cancer Res Treat        ISSN: 1533-0338


  1 in total

1.  Evaluating the Impact of Ionization Chamber-Specific Beam Quality Correction Factor in Dosimetry of Filtered and Unfiltered Photon Beams.

Authors:  Richa Sharma; Sunil Dutt Sharma; Priyanka Agarwal; Devesh Kumar Avasthi; Rohit Verma
Journal:  J Med Phys       Date:  2022-08-05
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.