Bioresorbable vascular scaffolds in saphenous vein grafts (data from OCTOPUS registry).

Tomasz Roleder1, Wojciech Wanha1, Grzegorz Smolka1, Jacek Zimoch1, Andrzej Ochała1, Wojciech Wojakowski1.   

Abstract

Entities:  

Year:  2015        PMID: 26677383      PMCID: PMC4679801          DOI: 10.5114/pwki.2015.55604

Source DB:  PubMed          Journal:  Postepy Kardiol Interwencyjnej        ISSN: 1734-9338            Impact factor:   1.426


× No keyword cloud information.

Introduction

Percutaneous coronary intervention (PCI) of saphenous vein grafts (SVG) with metallic stents is associated with a high rate of restenosis (19–21%) and target vessel failure in the long-term follow-up [1]. Potentially, implantation of a bioresorbable vascular scaffold (BVS) (ABSORB, Abbott Laboratories) into the stenotic SVG may be an alternative to metallic stents [2]. Although the effectiveness of BVS in treatment of de novo coronary lesions in native vessels is well documented [3], none of the previous studies evaluated the results of BVS implantation into SVG. Recently, we reported an imaging follow-up of a case of successful BVS implantation to treat a de novo stenotic SVG lesion [4-6]. Therefore, in this report we present pilot results from the OCTOPUS registry of a subgroup of patients with significant de novo SVG stenosis treated with BVS implantation.

Methods

OCTOPUS is a single-center registry evaluating the morphology of SVG lesions and stent healing by multimodality imaging including optical coherence imaging (OCT) and assessing long-term clinical outcomes of SVG interventions in patients with stable coronary artery disease (CAD) and acute coronary syndromes (ACS). Here we focus on patients treated with BVS implantation into SVG. The study was approved by the Ethics Committee of the Medical University of Silesia and conforms to the Declaration of Helsinki. All of the patients were enrolled in the study only after the patient gave their informed written consent.

Inclusion criteria

The registry included patients with a history of coronary artery bypass grafting utilizing SVG with recurrent stable CAD or ACS. The study exclusion criteria were as follows: age < 18 years old, glomerular filtration rate less then 45 ml/min/1.72 m2, severe valve disease warranting redo cardiac surgery and contrast allergy.

Quantitative coronary angiography

Quantitative angiography (QA) of SVG lesions was performed before and after BVS implantation using Horizon Cardiology 12.2 software (McKesson company) to evaluate reference lumen diameter (RLD), minimal lumen diameter (MLD), percentage diameter stenosis (%DS), and lesion length. At follow-up, MLD, %DS and late lumen loss (LLL) were evaluated. Late lumen loss was defined as the difference between the MLD immediately after the procedure and MLD at follow-up within the implanted BVS.

Optical coherence tomography imaging

Optical coherence tomography imaging was performed after the scaffold implantation and at the follow-up. The St Jude iLumien OPTIS Medical system was used for OCT imaging. The OCT probe (mid marker of the OCT Dragonfly catheter) was positioned 5 mm distally to the scaffold intended to be analyzed. All OCT imaging was performed using automated pullback triggered by the hand injection of contrast.

Optical coherence tomography analysis

CAAS IntraVascular software 1.1 (Pie Medical Imaging) was used for offline analysis of implanted BVS. The region of interest (ROI) was selected between proximal and distal edges of BVS where the OCT-visualized struts occupied more than 180° of the lumen's circumference. The analysis was performed every 1 mm to measure minimal lumen area (MLA) and minimal scaffold area (MSA). Minimal scaffold area was measured at the outer circumference of polymeric struts. The reference lumen area (RLA) was the average of the largest lumen area of two 5 mm long vessel segments located proximally and distally to the edges of BVS. Polymeric strut apposition was also assessed, and if there was a gap between the polymeric strut and the vessel's lumen contour, malapposition was diagnosed [7]. At the follow-up, polymeric strut coverage by neointima was also assessed. The complete coverage of BVS by neointima was identified if 4 corners of the polymeric strut had lost the right angle shape with signs of tissue coverage [8].

Statistical analysis

The statistical analysis was performed using MedCalc Software 14.12.0. As it was a small group of patients, the results were presented as the median and range.

Results

Patients’ characteristics

Six (4 male, 2 female) patients referred for PCI of de novo SVG lesions received BVS (Absorb, Abbott Laboratories). The median age was 73 years (range: 57–83 years). All patients had hypertension and 3 patients had diabetes. The median time from coronary artery bypass grafting to BVS implantation was 13 years (range: 3–22 years). All patients received dual antiplatelet therapy for at least 6 months or 12 months if presenting with ACS.

Procedural data

The median diameter of implanted BVS was 3.25 mm (range: 2.5–3.5 mm) with the median length 15 mm (range: 12–28 mm). For all BVS implantations distal protection devices were used and all treated lesions were predilated with a semi-/non-compliant (NC) balloon according to the operator's choice. For postdilation NC balloons were used.

Quantitative angiography analysis

Two BVS were implanted into the SVG of each right coronary artery, left anterior descending artery and circumflex artery. Quantitative angiography analysis revealed the median RLD of 3.13 mm (range: 2.25–3.45 mm) and median MLD of 1.12 mm (range: 0.52–1.86 mm) with median%DS = 61% (range: 46–72%). The 46% stenosis of one lesion was verified by OCT imaging, which revealed a 53% area stenosis with small MLA (RLA = 5.4 mm2, MLA = 2.5 mm2). The median lesion length was 14.4 mm (range: 6.6–20.4 mm). Post-procedure MLD of BVS was 2.59 mm (range: 2.0–3.0 mm) with median%DS = 13.5% (range: 8.1–16.7%). Four patients had angiography at 6-month follow-up and with median BVS MLD = 2.59 mm (range: 1.8–2.78 mm) and median%DS = 14.1% (range: 10.5–23%). Median LLL was 0.20 mm (range: 0.0–0.52 mm) at 6-month follow-up. No SVG restenosis was found at the site of BVS implantation at 6-month follow-up. One patient also had 18-month angiographic follow-up, which presented BVS MLD = 2.38 with LLL = 0.24 mm and%DS = 22%. Results of implantation were assessed by OCT in 3 BVS at baseline, in 3 BVS at 6-month follow-up, and 1 BVS 18 months after implantation. Two BVS had both OCT at baseline and OCT at 6-month follow-up.

Baseline

In total, 38 mm of BVS were analyzed at baseline. The RLA was 10.6 mm2, 9.1 mm2 and 5.4 mm2, and MSA was 7.7 mm2, 9.4 mm2 and 5.4 mm2, respectively. Twenty-three out of 151 polymeric struts (15%) were malapposed in the first patient (ROI = 11 mm) at baseline and 1 of 125 polymeric struts was malapposed (0.8%) in the second patient (ROI = 16 mm). The detected malappositions were not modified after BVS implantation. None of 111 polymeric struts was malapposed in the last patient (ROI = 11 mm). In one case a plaque protrusion and distal edge dissection were detected directly after implantation but without significant consequences (Figure 1). Thrombus was not found at baseline analysis.
Figure 1

Representative optical coherence tomography images of ABSORB implanted into saphenous vein graft. A – Malapposed polymeric struts (arrows). B – Polymeric struts not covered by neointima at 6-month follow-up (arrows). C – Plaque protrusion through polymeric struts (arrows). D – Vessel's dissection (white arrow) after BVS implantation with visible intimal flap (green arrow). E and F – Present OCT cross-section and longitudinal images of the same BVS at 6 and 18 months after implantation into right coronary saphenous vein graft. E – Two overhang polymeric struts covered by neointima (white arrows) 6 months after implantation. F – The same BVS 18 months after implantation with detected plaque protrusion at site of two overlapping polymeric struts

Representative optical coherence tomography images of ABSORB implanted into saphenous vein graft. A – Malapposed polymeric struts (arrows). B – Polymeric struts not covered by neointima at 6-month follow-up (arrows). C – Plaque protrusion through polymeric struts (arrows). D – Vessel's dissection (white arrow) after BVS implantation with visible intimal flap (green arrow). E and F – Present OCT cross-section and longitudinal images of the same BVS at 6 and 18 months after implantation into right coronary saphenous vein graft. E – Two overhang polymeric struts covered by neointima (white arrows) 6 months after implantation. F – The same BVS 18 months after implantation with detected plaque protrusion at site of two overlapping polymeric struts

6-month follow-up

Optical coherence tomography imaging of BVS at 6-month follow-up was performed in 3 patients with ROI = 18 mm, ROI = 12 mm and ROI = 22 mm at median 7 (5–7) months after the index procedure. The MSA at follow-up was 8.6 mm2, 5.7 mm2 and 8.2 mm2, respectively. All BVS analyzed at 6-month follow-up were well apposed. In the 2 first patients from baseline analysis 86% (153/177), and 87% (99/125) of polymeric struts were covered by neointima respectively. In the BVS of the other patient, that was not analyzed at baseline, 88% (25/202) of polymeric struts were covered by neointima at 6-month follow-up, even at the site of scaffold disruption (Figure 1).

18-month follow-up

Optical coherence tomography imaging at 18-month follow-up after implantation of BVS was performed only in 1 patient and showed that 97% (5/175) of polymeric struts were covered by neointima with MLA 5.2 mm2 and MSA 8.3 mm2. At the site of BVS disruption that was identified 6 months after its implantation, a lipid-rich plaque protrusion was detected at 18-month follow-up (Figure 1).

Patients’ follow-up

All patients were followed up to 12 months after BVS implantation. There were no deaths or target vessel revascularization defined as any revascularization within the treated vessel.

Discussion

Implantation of BVS is safe and feasible providing the diameter of the scaffold falls within the recommended limit of scaffold expansion. Therefore this approach is limited to conduits up to 4 mm in diameter. This report for the first time provides the OCT analysis of BVS healing in SVG. At the time of implantation, a similar ratio of malapposed polymeric struts was observed as in native coronary arteries, as previously reported [8]. However, the BVS coverage was less evident as compared to that observed in native coronary arteries at follow-up [7]. It is probably a similar phenomenon as in the metallic everolimus-eluting stent implanted into the SVG [9], which suggested that the degree of healing might be related to the particular anatomy and tissue properties of SVG. Moreover, the presence of atheroma that protruded through polymeric struts 18 months after implantation suggested that either neoatherosclerosis developed or the mechanical properties of the graft changed the geometry of the ROI [10]. It warrants further large clinical studies to investigate the safety and efficacy of ABSORB implantation into the SVG.

Conclusions

This report presents long-term imaging data on a small number of patients with SVG stenosis treated with BVS. It suggests that the healing of BVS in SVG is not impaired, but a properly sized clinical trial is needed to confirm this.
  10 in total

1.  Serial analysis of the malapposed and uncovered struts of the new generation of everolimus-eluting bioresorbable scaffold with optical coherence tomography.

Authors:  Josep Gomez-Lara; Maria Radu; Salvatore Brugaletta; Vasim Farooq; Roberto Diletti; Yoshinobu Onuma; Stephan Windecker; Leif Thuesen; Dougal McClean; Jacques Koolen; Robert Whitbourn; Dariusz Dudek; Pieter C Smits; Evelyn Regar; Susan Veldhof; Richard Rapoza; John A Ormiston; Hector M Garcia-Garcia; Patrick W Serruys
Journal:  JACC Cardiovasc Interv       Date:  2011-09       Impact factor: 11.195

2.  Increased thin-cap neoatheroma and periprocedural myocardial infarction in drug-eluting stent restenosis: multimodality intravascular imaging of drug-eluting and bare-metal stents.

Authors:  Ziad A Ali; Tomasz Roleder; Jagat Narula; Bibhu D Mohanty; Usman Baber; Jason C Kovacic; Gary S Mintz; Fumiyuki Otsuka; Stephen Pan; Renu Virmani; Samin K Sharma; Pedro Moreno; Annapoorna S Kini
Journal:  Circ Cardiovasc Interv       Date:  2013-09-24       Impact factor: 6.546

3.  Late outcomes of drug eluting and bare metal stents in saphenous vein graft percutaneous coronary intervention.

Authors:  Satheesh Nair; Farzin Fath-Ordoubadi; Bernard Clarke; Magdi El-Omar; James Foley; Doug G Fraser; Vaikom S Mahadevan; Ludwig Neyses; Raj S Khattar; Mamas A Mamas
Journal:  EuroIntervention       Date:  2011-03       Impact factor: 6.534

4.  Bioresorbable vascular scaffolds in patients with acute coronary syndromes: the POLAR ACS study.

Authors:  Dariusz Dudek; Łukasz Rzeszutko; Wojciech Zasada; Rafał Depukat; Zbigniew Siudak; Andrzej Ochała; Wojciech Wojakowski; Tadeusz Przewłocki; Krzysztof Żmudka; Janusz Kochman; Andrzej Lekston; Mariusz Gąsior
Journal:  Pol Arch Med Wewn       Date:  2014

5.  "First-in-man" use of bioresorbable vascular scaffold in saphenous vein graft.

Authors:  Paul Jau Lueng Ong; Fahim Haider Jafary; Hee Hwa Ho
Journal:  EuroIntervention       Date:  2013-05-20       Impact factor: 6.534

6.  Implantation of a bioabsorbable vascular scaffold into a coronary vein graft: a two-week angiography follow-up.

Authors:  Tomasz Roleder; Andrzej Ochała; Grzegorz Smolka; Wojciech Wańha; Wojciech Wojakowski
Journal:  Kardiol Pol       Date:  2014       Impact factor: 3.108

7.  Optical coherence tomography imaging of everolimus-eluting bioresorbable vascular scaffold implanted into coronary vein graft at 3-month follow-up.

Authors:  Tomasz Roleder; Zofia Parma; Grzegorz Smolka; Andrzej Ochała; Wojciech Wojakowski
Journal:  Eur Heart J       Date:  2014-04-10       Impact factor: 29.983

8.  Differences in neointimal thickness between the adluminal and the abluminal sides of malapposed and side-branch struts in a polylactide bioresorbable scaffold: evidence in vivo about the abluminal healing process.

Authors:  Juan Luis Gutiérrez-Chico; Frank Gijsen; Evelyn Regar; Jolanda Wentzel; Bernard de Bruyne; Leif Thuesen; John Ormiston; Dougal R McClean; Stephan Windecker; Bernard Chevalier; Dariusz Dudek; Robert Whitbourn; Salvatore Brugaletta; Yoshinobu Onuma; Patrick W Serruys
Journal:  JACC Cardiovasc Interv       Date:  2012-04       Impact factor: 11.195

9.  Optical coherence tomography analysis of the stenting of saphenous vein graft (SOS) Xience V Study: use of the everolimus-eluting stent in saphenous vein graft lesions.

Authors:  Aristotelis C Papayannis; Tesfaldet T Michael; Dinara Yangirova; Abdul-Rahman Abdel-Karim; Julia Kohlhaas; Arif Mahmood; Tayo Addo; Donald Haagen; Lorenza Makke; Michele Roesle; Bavana Rangan; Subhash Banerjee; Emmanouil S Brilakis
Journal:  J Invasive Cardiol       Date:  2012-08       Impact factor: 2.022

Review 10.  "Leaving nothing behind": is the bioresorbable vascular scaffold a new hope for patients with coronary artery disease?

Authors:  Maciej Lesiak; Aleksander Araszkiewicz
Journal:  Postepy Kardiol Interwencyjnej       Date:  2014-11-17       Impact factor: 1.426

  10 in total
  5 in total

1.  Bioresorbable scaffolds: should we stay simple or go complex?

Authors:  Luis Ortega-Paz; Salvatore Brugaletta; Hector M Garcia-Garcia; Manel Sabaté
Journal:  Cardiovasc Diagn Ther       Date:  2017-06

2.  Off-label indications for bioresorbable scaffolds: "Beethoven can, but you cannot".

Authors:  Juan Luis Gutiérrez-Chico; Undine Ella Witt; Milosz Jaguszewski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2016-02-11       Impact factor: 1.426

3.  Saphenous graft atherosclerosis as assessed by optical coherence tomography data for stenotic and non-stenotic lesions from the OCTOPUS registry.

Authors:  Grzegorz M Kubiak; Elżbieta Pociask; Wojciech Wańha; Magdalena Dobrolińska; Paweł Gąsior; Grzegorz Smolka; Andrzej Ochała; Zbigniew Gąsior; Wojciech Wojakowski; Tomasz Roleder
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-06-19       Impact factor: 1.426

4.  Prediction models for different plaque morphology in non-significantly stenosed regions of saphenous vein grafts assessed with optical coherence tomography.

Authors:  Grzegorz M Kubiak; Magdalena Dobrolińska; Elżbieta Pociask; Wojciech Wańha; Paweł Gąsior; Grzegorz Smolka; Andrzej Ochała; Wojciech Wojakowski; Tomasz Roleder
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-12-11       Impact factor: 1.426

5.  Multimodality intravascular imaging of bioresorbable vascular scaffolds implanted in vein grafts.

Authors:  Tomasz Roleder; Elzbieta Pociask; Wojciech Wanha; Pawel Gasior; Magdalena Dobrolinska; Magdalena Garncarek; Przemyslaw Pietraszewski; Radoslaw Kurzelowski; Grzegorz Smolka; Wojciech Wojakowski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2019-06-26       Impact factor: 1.426

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.