INTRODUCTION: Human metapneumovirus (hMPV) is an emergent human respiratory pathogen. This study aimed to evaluate the performance of direct immunofluorescence (DIF) to detect hMPV in a clinical laboratory setting. METHODS: Nasopharyngeal aspirate samples (448) of children and adults with respiratory illness were used to detect hMPV by using DIF and real time quantitative reverse transcription-polymerase chain reaction (qRT-PCR) assays. RESULTS: In all, 36 (8%) samples were positive by DIF and 94 (21%) were positive by qRT-PCR. Direct immunofluorescence specificity was 99% and sensitivity was 38%. CONCLUSIONS: DIF is not very sensitive under clinical laboratory settings.
INTRODUCTION:Human metapneumovirus (hMPV) is an emergent human respiratory pathogen. This study aimed to evaluate the performance of direct immunofluorescence (DIF) to detect hMPV in a clinical laboratory setting. METHODS: Nasopharyngeal aspirate samples (448) of children and adults with respiratory illness were used to detect hMPV by using DIF and real time quantitative reverse transcription-polymerase chain reaction (qRT-PCR) assays. RESULTS: In all, 36 (8%) samples were positive by DIF and 94 (21%) were positive by qRT-PCR. Direct immunofluorescence specificity was 99% and sensitivity was 38%. CONCLUSIONS: DIF is not very sensitive under clinical laboratory settings.
Authors: Pamela Elizabeth Rodriguez; María Celia Frutos; María Pilar Adamo; Cecilia Cuffini; Jorge Augusto Cámara; María Gabriela Paglini; Laura Moreno; Alicia Cámara Journal: PLoS One Date: 2020-12-28 Impact factor: 3.240