Literature DB >> 26663544

Inter-Observer, Intra-Observer and Intra-Individual Reliability of Uroflowmetry Tests in Aged Men: A Generalizability Theory Approach.

Ying-Buh Liu1, Stephen S Yang1, Cheng-Hsing Hsieh1, Chia-Da Lin1, Shang-Jen Chang1.   

Abstract

OBJECTIVES: To evaluate the inter-observer, intra-observer and intra-individual reliability of uroflowmetry and post-void residual urine (PVR) tests in adult men.
METHODS: Healthy volunteers aged over 40 years were enrolled. Every participant underwent two sets of uroflowmetry and PVR tests with a 2-week interval between the tests. The uroflowmetry tests were interpreted by four urologists independently. Uroflowmetry curves were classified as bell-shaped, bell-shaped with tail, obstructive, restrictive, staccato, interrupted and tower-shaped and scored from 1 (highly abnormal) to 5 (absolutely normal). The agreements between the observers, interpretations and tests within individuals were analyzed using kappa statistics and intraclass correlation coefficients. Generalizability theory with decision analysis was used to determine how many observers, tests, and interpretations were needed to obtain an acceptable reliability (> 0.80).
RESULTS: Of 108 volunteers, we randomly selected the uroflowmetry results from 25 participants for the evaluation of reliability. The mean age of the studied adults was 55.3 years. The intra-individual and intra-observer reliability on uroflowmetry tests ranged from good to very good. However, the inter-observer reliability on normalcy and specific type of flow pattern were relatively lower. In generalizability theory, three observers were needed to obtain an acceptable reliability on normalcy of uroflow pattern if the patient underwent uroflowmetry tests twice with one observation.
CONCLUSIONS: The intra-individual and intra-observer reliability on uroflowmetry tests were good while the inter-observer reliability was relatively lower. To improve inter-observer reliability, the definition of uroflowmetry should be clarified by the International Continence Society.
© 2013 Wiley Publishing Asia Pty Ltd.

Entities:  

Keywords:  inter-observer reliability; intra-individual reliability; intra-observer reliability; uroflowmetry

Year:  2013        PMID: 26663544     DOI: 10.1111/luts.12022

Source DB:  PubMed          Journal:  Low Urin Tract Symptoms        ISSN: 1757-5664            Impact factor:   1.592


  2 in total

1.  A comparison of flow rate curve shape and video-urodynamic findings in women with lower urinary tract symptoms: can flow rate curve shape predict female bladder outflow obstruction or detrusor underactivity?

Authors:  Huriye Kocadag; Bogdan Toia; Richard Axell; Habiba Yasmin; Mahreen H Pakzad; Rizwan Hamid; Tamsin J Greenwell; Jeremy L Ockrim
Journal:  World J Urol       Date:  2020-07-11       Impact factor: 4.226

2.  Development of an Automatic Interpretation Algorithm for Uroflowmetry Results: Application of Artificial Intelligence.

Authors:  Min Soo Choo; Ho Young Ryu; Sangchul Lee
Journal:  Int Neurourol J       Date:  2022-03-31       Impact factor: 2.835

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.