K Mitev1, P Cassette2, S Georgiev3, I Dimitrova3, B Sabot2, T Boshkova3, I Tartès2, D Pressyanov3.
Abstract
This work demonstrates that a Liquid Scintillation Counting (LSC) technique using a Triple to Double Coincidence Ratio counter with extending dead-time is very appropriate for the accurate measurement of (222)Rn activity absorbed in thin polycarbonate foils. It is demonstrated that using a toluene-based LS cocktail, which dissolves polycarbonates, the (222)Rn activity absorbed in thin Makrofol N foil can be determined with a relative standard uncertainty of about 0.7%. A LSC-based application of the methodology for determination of the diffusion length of (222)Rn in thin polycarbonate foils is proposed and the diffusion length of (222)Rn in Makrofol N (38.9±1.3µm) and the partition coefficient of (222)Rn in Makrofol N from air (112±12, at 20°C) and from water (272±17, at 21°C) are determined. Calibration of commercial LS spectrometers for (222)Rn measurements by LSC of thin polycarbonate foils is performed and the minimum detectable activities by this technique are estimated.Entities:
Keywords: (222)Rn; LSC; Polycarbonate foil; TDCR
Year: 2015 PMID: 26640234 DOI: 10.1016/j.apradiso.2015.11.047
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513