Literature DB >> 26634970

Experimental and simulation investigation of ion transfer in different sampling capillaries.

Quan Yu1, Tao Jiang1, Kai Ni1, Xiang Qian1, Fei Tang2, Xiaohao Wang1,2.   

Abstract

Atmospheric pressure interfaces were a fundamental structure for transferring air generated ions into the vacuum manifold of a mass spectrometer. This work is devoted to the characterization of ion transfer in metal capillaries through both experimental and simulated investigations. The impact of capillary configurations on ion transmission efficiency was evaluated using an electrospray mass spectrometer with various bent capillaries as the transfer devices. In addition, a numerical model has been set up by coupling the SIMION 8.0 and the computational flow dynamics for simulation study of ion migration in the complex atmospheric system. The transfer efficiency was found to be highly affected by the variation in electric field and the capillary geometry, revealing that the hydrodynamic and electric force were both dominant and interactional during the transmission process. The consistency of the results from the experimental analysis and simulation modeling proved the validity of the model, which was helpful for understanding ion activity in transfer capillaries.
Copyright © 2015 John Wiley & Sons, Ltd.

Keywords:  atmospheric pressure; gas flow; ion transfer; sampling capillary; simulation modeling

Year:  2015        PMID: 26634970     DOI: 10.1002/jms.3707

Source DB:  PubMed          Journal:  J Mass Spectrom        ISSN: 1076-5174            Impact factor:   1.982


  1 in total

1.  Gas Flow in the Capillary of the Atmosphere-to-Vacuum Interface of Mass Spectrometers.

Authors:  Michael Skoblin; Alexey Chudinov; Ilia Soulimenkov; Vladimir Brusov; Viacheslav Kozlovskiy
Journal:  J Am Soc Mass Spectrom       Date:  2017-07-18       Impact factor: 3.109