Literature DB >> 26632858

Comparison of the transcutaneous bilirubinometers BiliCare and Minolta JM-103 in late preterm and term neonates.

Simone Pratesi1, Luca Boni2,3, Lorenzo Tofani2,3, Elettra Berti1, Sara Sollai1, Carlo Dani4.   

Abstract

OBJECTIVES: To assess the agreement of transcutaneous bilirubin (TcB) measurement with the Bilicare™ System in comparison to TcB measured with JM-103™ and total serum bilirubin (TSB).
METHODS: Caucasian infants with gestational age ≥35 weeks with non-hemolytic jaundice received TcB measurement with both Bilicare™ and JM-103™ devices. TSB was also obtained in infants at risk of phototherapy.
RESULTS: We studied 458 infants measuring TcB with Bilicare™ and JM-103™, correlating the results and with TSB. The mean difference ± 2SD between Bilicare™ and JM-103™ TcB was 2.02 ± 4.46 mg/dL and decreased from 2.88 ± 3.17 to 1.20 ± 4.55, and to -0.95 ± 4.58 mg/dL at mild, moderate and high TcB values, respectively.
CONCLUSIONS: Bilicare™ and JM-103™ TcB measurements are well correlated, but Bilicare™ over-estimates TcB for mild and moderate values and under-estimates it for high values compared to JM-103™. This could increase the prescription of TSB measurements for less serious cases and decrease them in the most worrisome.

Entities:  

Keywords:  Infant; jaundice; transcutaneous bilirubinometry

Mesh:

Substances:

Year:  2015        PMID: 26632858     DOI: 10.3109/14767058.2015.1113521

Source DB:  PubMed          Journal:  J Matern Fetal Neonatal Med        ISSN: 1476-4954


  5 in total

1.  Israel transcutaneous bilirubin nomogram predicts significant hyperbilirubinemia.

Authors:  R Bromiker; A Goldberg; M Kaplan
Journal:  J Perinatol       Date:  2017-08-31       Impact factor: 2.521

2.  Accuracy of enhanced transcutaneous bilirubinometry according to various measurement sites.

Authors:  Lucia Casnocha Lucanova; Jana Zibolenova; Katarina Matasova; Lenka Docekalova; Mirko Zibolen
Journal:  Turk Arch Pediatr       Date:  2021-01-01

3.  Large scale validation of a new non-invasive and non-contact bilirubinometer in neonates with risk factors.

Authors:  Animesh Halder; Aniruddha Adhikari; Ria Ghosh; Soumendra Singh; Amrita Banerjee; Nilanjana Ghosh; Arnab Madhab Bhattacharya; Shrabani Mandal; Prantar Chakrabarti; Debasis Bhattacharyya; Hatem M Altass; Moataz Morad; Saleh A Ahmed; Asim Kumar Mallick; Samir Kumar Pal
Journal:  Sci Rep       Date:  2020-07-07       Impact factor: 4.379

4.  Reproducibility of BiliCare™ Transcutaneus Bilirrubin Meter in Mexican Newborns.

Authors:  David Oldak; Gabriela García; Estela E Gonzalez; Enver Aillon; Juan C Falcon; Ernesto Ayala; Bernardo Oldak
Journal:  Int J Pediatr       Date:  2019-01-01

5.  Transcutaneous bilirubin measurement in healthy Saudi term newborns.

Authors:  Mahmood D Al-Mendalawi
Journal:  Saudi Med J       Date:  2016-05       Impact factor: 1.484

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.