BrachyView, a novel in-body imaging system for HDR prostate brachytherapy: Experimental evaluation.

M Safavi-Naeini1, Z Han1, S Alnaghy1, D Cutajar1, M Petasecca1, M L F Lerch1, D R Franklin2, J Bucci3, M Carrara4, M Zaider5, A B Rosenfeld1.   

Abstract

PURPOSE: This paper presents initial experimental results from a prototype of high dose rate (HDR) BrachyView, a novel in-body source tracking system for HDR brachytherapy based on a multipinhole tungsten collimator and a high resolution pixellated silicon detector array. The probe and its associated position estimation algorithms are validated and a comprehensive evaluation of the accuracy of its position estimation capabilities is presented.
METHODS: The HDR brachytherapy source is moved through a sequence of positions in a prostate phantom, for various displacements in x, y, and z. For each position, multiple image acquisitions are performed, and source positions are reconstructed. Error estimates in each dimension are calculated at each source position and combined to calculate overall positioning errors. Gafchromic film is used to validate the accuracy of source placement within the phantom.
RESULTS: More than 90% of evaluated source positions were estimated with an error of less than one millimeter, with the worst-case error being 1.3 mm. Experimental results were in close agreement with previously published Monte Carlo simulation results.
CONCLUSIONS: The prototype of HDR BrachyView demonstrates a satisfactory level of accuracy in its source position estimation, and additional improvements are achievable with further refinement of HDR BrachyView's image processing algorithms.

Mesh:

Substances:

Year:  2015        PMID: 26632063     DOI: 10.1118/1.4935866

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Preliminary Monte Carlo Investigation of Using Ir-192 as the Source for Real Time Imaging Purpose.

Authors:  Chengyu Shi; Brian Wang
Journal:  Int J Med Phys Clin Eng Radiat Oncol       Date:  2017-01-12

2.  Small-field dosimetry with detector-specific output correction factor for single-isocenter stereotactic radiotherapy of single and multiple brain metastases.

Authors:  Tomohiro Ono; Kohei Kawata; Mitsuhiro Nakamura; Megumi Uto; Takashi Mizowaki
Journal:  Radiol Phys Technol       Date:  2022-10-22

Review 3.  In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Gabriel P Fonseca; Jacob G Johansen; Ryan L Smith; Luc Beaulieu; Sam Beddar; Gustavo Kertzscher; Frank Verhaegen; Kari Tanderup
Journal:  Phys Imaging Radiat Oncol       Date:  2020-09-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.