Janine Pilcher1, Philippa Shirtcliffe1, Mitesh Patel2, Steve McKinstry3, Terrianne Cripps1, Mark Weatherall4, Richard Beasley1.
Abstract
BACKGROUND: Electronic monitoring of inhaled asthma therapy is suggested as the 'gold standard' for measuring patterns of medication use in clinical trials. The SmartTurbo (Adherium (NZ) Ltd, Auckland, New Zealand) is an electronic monitor for use with a turbuhaler device (AstraZeneca, UK). The aim of this study was to determine the accuracy of the SmartTurbo in recording Symbicort actuations over a 12-week period of use.Entities:
Keywords: Asthma
Year: 2015 PMID: 26629345 PMCID: PMC4653861 DOI: 10.1136/bmjresp-2015-000097
Source DB: PubMed Journal: BMJ Open Respir Res ISSN: 2052-4439
Figure 1SmartTurbo monitor.
Figure 2SmartTurbo monitor attached to Symbicort turbuhaler.
Monitor function testing over 84 days
| Action | Day 0 | Days 5, 6, 8, 9 | Days 7, 14 and 21 | Day 28 | Day 56 | Day 84 |
|---|---|---|---|---|---|---|
| Attachment of SmartTurbo to a Symbicort turbuhaler prior to any inhaler actuations | X | X* | X* | |||
| Initial screen of monitor | X | |||||
| Low-use actuations | X | X | X | X | X | X |
| High-use actuations | X | X | X | X | ||
| Storage of electronic data for 28-day period | X | |||||
| Upload of data to website following and on the same day as inhaler actuations | X | X | X | X | X | |
| Upload of data obtained on day 56 | X | |||||
| Comparison of uploaded website data to diary data† | X | X | X | X | X | |
| Comparison of website data to diary data time | X | X | X | X | X | |
| Monitor battery charge check | X |
Testing may occur ±1 day.
*Previous inhaler removed on same day.
†Recording of any missed actuations or spurious actuations.
Overall results from the testing process
| Monitor electronically recorded actuation | Written diary actuation log | |
|---|---|---|
| Yes | No | |
| Yes | 2796 | 3 |
| No | 4 | |
| Sensitivity* 99.9% (99.6%) | ||
1. Data from day 0 uploaded to V.7.7 Smartinhaler Connection Centre programme. Subsequently, all data were uploaded to V.7.8 Smartinhaler Connection Centre programme.
2. Investigators were unable to access the Adherium website to upload data on day 14, so data were uploaded on day 15 when access was gained.
3. All data from day 14 (320/320 actuations) were 1 h discrepant (to within 3 min), in keeping with the time prior to the cessation of the New Zealand daylight savings period on day 8. This discrepancy applied to data on day 14 only.
*Sensitivity is the proportion of diary actuations electronically recorded by the monitor expressed as a percentage (lower 97.5% confidence bound). Positive predictive value is the proportion of monitor recorded actuations that were recorded in the diary expressed as a percentage (lower 97.5% confidence bound).
Detailed results from the testing process
| Monitor function check | Outcome | |
|---|---|---|
| Monitors passed initial screening test | 20/20 (100%) | |
| Sensitivity and positive predictive value by factors potentially affecting accuracy | ||
| Regimen | ||
| Low-use actuations | 878/880 (99.8) (99.2) | 878/881 (99.7) (99.0) |
| High-use actuations | 1918/1920 (99.9) (99.6) | 1918/1918 (100) (99.8) |
| Period, days | ||
| 0 (screening) | 80/80 (100) (95.5) | 80/83 (96.4) (89.8) |
| 1–28 | 1918/1920 (99.9) (99.6) | 1918/1918 (100) (99.8) |
| 29–56 | 399/400 (99.8) (98.6) | 399/399 (100) (99.1) |
| 57–84 | 399/400 (99.8) (98.6) | 399/399 (100) (99.1) |
| Upload pattern | ||
| Same day upload | 2397/2400 (99.9) (99.6) | 2397/2400 (99.9) (99.6) |
| Delayed upload (days 56–84) | 399/400 (99.8) (98.6) | 399/399 (100) (99.1) |
| Turbuhaler attached to monitor prior to testing | ||
| No | 1917/1920 (99.8) (99.5) | 1917/1917 (100) (99.8) |
| Yes (days 0, 28 and 56) | 879/880 (99.9) (99.4) | 879/882 (99.7) (99.0) |
| Accuracy by monitor** | ||
| Number of monitors with at least one missed actuation recording | 3/20 (15) | |
| Number of monitors with at least one spurious actuation recording | 2/20 (10) | |
| Monitor battery check** | ||
| Pass on day 84 | 20/20 (100) | |
1. Data from day 0 uploaded to V.7.7 Smartinhaler Connection Centre programme. Subsequently, all data were uploaded to V.7.8 Smartinhaler Connection Centre programme.
2. Investigators were unable to access the Adherium website to upload data on day 14, so data were uploaded on day 15 when access was gained.
3. All data from day 14 (320/320 actuations) were 1 h discrepant (to within 3 min), in keeping with the time prior to the cessation of the New Zealand daylight savings period on day 8. This discrepancy applied to data on day 14 only.
*Sensitivity is the proportion of diary actuations recorded by the monitor expressed as N/N (%) (lower 97.5% confidence bound). Positive predictive value is the proportion of monitor recorded actuations that were recorded in the diary expressed as N/N (%) (lower 97.5% confidence bound).
**N/N (%) (95% CI).