Literature DB >> 26628564

Comparison of Transcutaneous and Capillary Measurement of PCO2 in Hypercapnic Subjects.

Sven Stieglitz1, Sandhya Matthes2, Christina Priegnitz2, Lars Hagmeyer2, Winfried Randerath2.   

Abstract

BACKGROUND: Measurement of PCO2 is vital in determining effective alveolar ventilation. However, obtaining capillary PCO2 by a skin prick of the earlobe is painful, and nocturnal measurements disturb sleep. End-expiratory measurement of PCO2 is also well established, but there is a low precision in predicting arterial or capillary CO2. The purpose of the study was to evaluate nocturnal measurement of noninvasive, transcutaneous PCO2 (PtcCO2 ) measurement in hypercapnic subjects.
METHODS: In this prospective study, 31 subjects with chronic hypercapnic failure--in a stable phase of the underlying disease--and a control group of 12 healthy volunteers were included. Transcutaneous measurements were taken by the Tosca sensor (Radiometer, Copenhagen, Denmark) over a period of at least 6 h during the night. A capillary blood gas was measured at midnight and 4:00 am.
RESULTS: The mean nocturnal capillary PCO2 (PcapCO2 ) of subjects was 50.6 ± 10.2 mm Hg. In the 31 subjects with known hypercapnic respiratory failure, the correlation between PtcCO2 and PcapCO2 at midnight was 0.86 and at 4:00 am r = 0.80. The bias of the hypercapnic subjects was d = + 4.5 with a limit(s) of agreement of 2 SD = 13.0. The process of blood sampling caused no significant change in PtcCO2 .
CONCLUSIONS: Our study evaluated transcutaneous capnography as a continuous nocturnal measurement in hypercapnic subjects. We found a good agreement between the methods. Because CO2 is not constant in patients with respiratory failure, but instead fluctuates, we would recommend the continuous transcutaneous measurement of PCO2 as our method of choice in the diagnosis of nocturnal hypercapnia.
Copyright © 2016 by Daedalus Enterprises.

Entities:  

Keywords:  chronic obstructive pulmonary disease; hypercapnic failure; hypoventilation; noninvasive measurement of carbon dioxide pressure; obesity-hypoventilation syndrome; sleep-disordered breathing

Mesh:

Substances:

Year:  2015        PMID: 26628564     DOI: 10.4187/respcare.03917

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  3 in total

1.  Acute Responses to Oxygen Delivery via High Flow Nasal Cannula in Patients with Severe Chronic Obstructive Pulmonary Disease-HFNC and Severe COPD.

Authors:  Amy H Attaway; Jihane Faress; Frank Jacono; Srinivasan Dasarathy
Journal:  J Clin Med       Date:  2021-04-21       Impact factor: 4.241

2.  Evaluation of time courses of agreement between minutely obtained transcutaneous blood gas data and the gold standard arterial data from spontaneously breathing Asian adults, and various subgroup analyses.

Authors:  Akira Umeda; Masahiro Ishizaka; Masamichi Tasaki; Tateki Yamane; Taiji Watanabe; Yasushi Inoue; Taichi Mochizuki; Yasumasa Okada; Sarah Kesler
Journal:  BMC Pulm Med       Date:  2020-05-29       Impact factor: 3.317

3.  Transcutaneous partial pressure of carbon dioxide monitoring during EUS-guided drainage of peripancreatic fluid collections using carbon dioxide insufflation: A prospective study.

Authors:  Liu Xiang; Sun Jiayi; Wang Guoxin; Ge Nan; Wang Sheng; Guo Jintao; Sun Siyu
Journal:  Endosc Ultrasound       Date:  2020 Jan-Feb       Impact factor: 5.628

  3 in total