Assessment of cardiac mass from tagged magnetic resonance images.

Abram W Makram1, Ayman M Khalifa2, Hossam El-Rewaidy3, Ahmed S Fahmy4, El-Sayed H Ibrahim5.   

Abstract

PURPOSE: Tagged and cine magnetic resonance imaging (tMRI and cMRI) techniques are used for evaluating regional and global heart function, respectively. Measuring global function parameters directly from tMRI is challenging due to the obstruction of the anatomical structure by the tagging pattern. The purpose of this study was to develop a method for processing the tMRI images to improve the myocardium-blood contrast in order to estimate global function parameters from the processed images.
MATERIALS AND METHODS: The developed method consists of two stages: (1) removing the tagging pattern based on analyzing and modeling the signal distribution in the image's k-space, and (2) enhancing the blood-myocardium contrast based on analyzing the signal intensity variability in the two tissues. The developed method is implemented on images from twelve human subjects.
RESULTS: Ventricular mass measured with the developed method showed good agreement with that measured from gold-standard cMRI images. Further, preliminary results on measuring ventricular volume using the developed method are presented.
CONCLUSION: The promising results in this study show the potential of the developed method for evaluating both regional and global heart function from a single set of tMRI images, with associated reduction in scan time and patient discomfort.

Entities:  

Keywords:  Cardiac mass; Detagging; MRI; Mean shift theorem; Tagging

Mesh:

Substances:

Year:  2015        PMID: 26627894     DOI: 10.1007/s11604-015-0504-4

Source DB:  PubMed          Journal:  Jpn J Radiol        ISSN: 1867-1071            Impact factor:   2.374


  3 in total

1.  Improved accuracy of quantitative assessment of left ventricular volume and ejection fraction by geometric models with steady-state free precession.

Authors:  Holger Thiele; Ingo Paetsch; Bernhard Schnackenburg; Axel Bornstedt; Olaf Grebe; Ernst Wellnhofer; Gerhard Schuler; Eckart Fleck; Eike Nagel
Journal:  J Cardiovasc Magn Reson       Date:  2002       Impact factor: 5.364

2.  Comparison between tagged MRI and standard cine MRI for evaluation of left ventricular ejection fraction.

Authors:  Christophe Dornier; G Aernout Somsen; Marko K Ivancevic; Nael F Osman; Dominique Didier; Alberto Righetti; Jean-Paul Vallée
Journal:  Eur Radiol       Date:  2004-04-06       Impact factor: 5.315

3.  Accurate estimation of the myocardium global function from reduced magnetic resonance image acquisitions.

Authors:  Hossam El-Rewaidy; Ayman Khalifa; Ahmed S Fahmy
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2014
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.