Literature DB >> 2662554

A 20-MHz ultrasound system for imaging the intestinal wall.

R W Martin1, F E Silverstein, M B Kimmey.   

Abstract

An ultrasound system has been developed which uses high-frequency (20 MHz) ultrasound to provide high-resolution images of tissue. The system provides 0.21-mm range and 0.65-mm lateral resolution. The transducer aperture size is 1.8 mm maximum. Miniature probes have been developed which can image via the biopsy channels of standard fiberoptic endoscopes as well as probes for imaging in vitro. A commercially available video "frame grabber" is used in conjunction with a standard microcomputer for image acquisition. This allows images to be displayed and recorded on standard television equipment and be stored and manipulated digitally. The features of the system allow in vivo imaging, in vitro imaging after resection, and histological images of the same tissue region to be acquired and compared. This method is particularly useful in learning how to correctly interpret ultrasonic images of the intestinal wall. The use of 20 MHz is advantageous in achieving excellent resolution and small size probes. The system provides a unique approach to imaging the intestinal wall.

Mesh:

Year:  1989        PMID: 2662554     DOI: 10.1016/0301-5629(89)90071-9

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  3 in total

Review 1.  Design characteristics for intravascular ultrasonic catheters.

Authors:  R W Martin; C C Johnson
Journal:  Int J Card Imaging       Date:  1989

Review 2.  Endoscopic ultrasound advances, part 1: diagnosis.

Authors:  Edward Kim; Jennifer J Telford
Journal:  Can J Gastroenterol       Date:  2009-09       Impact factor: 3.522

3.  High-frequency ultrasound probe sonography staging for colorectal neoplasia with superficial morphology: its utility and impact on patient management.

Authors:  Ondrej Urban; Martin Kliment; Petr Fojtik; Premysl Falt; Julius Orhalmi; Petr Vitek; Pavol Holeczy
Journal:  Surg Endosc       Date:  2011-05-18       Impact factor: 4.584