Literature DB >> 26594097

Ethylene Oxide and Hydrogen Peroxide Gas Plasma Sterilization: Precautionary Practices in U.S. Hospitals.

James M Boiano1, Andrea L Steege1.   

Abstract

OBJECTIVE: Evaluate precautionary practices and extent of use of ethylene oxide (EtO) and hydrogen peroxide gas plasma (HPGP) sterilization systems, including use of single chamber EtO units.
DESIGN: Modular, web-based survey. PARTICIPANTS: Members of professional practice organizations who reported using EtO or HPGP in the past week to sterilize medical instruments and supplies. Participating organizations invited members via email which included a hyperlink to the survey.
METHODS: Descriptive analyses were conducted including simple frequencies and prevalences.
RESULTS: A total of 428 respondents completed the module on chemical sterilants. Because most respondents worked in hospitals (87%, n=373) analysis focused on these workers. Most used HPGP sterilizers (84%, n=373), 38% used EtO sterilizers, with 22% using both. Nearly all respondents using EtO operated single chamber units (94%, n=120); most of them reported that the units employed single use cartridges (83%, n=115). Examples of where engineering and administrative controls were lacking for EtO include: operational local exhaust ventilation (7%; n=114); continuous air monitoring (6%; n=113); safe handling training (6%; n=142); and standard operating procedures (4%; n=142). Examples of practices which may increase HPGP exposure risk included lack of standard operating procedures (9%; n=311) and safe handling training (8%; n=312).
CONCLUSIONS: Use of precautionary practices was good but not universal. EtO use appears to have diminished in favor of HPGP which affords higher throughput and minimal regulatory constraints. Separate EtO sterilization and aeration units were still being used nearly one year after U.S. EPA prohibited their use.

Entities:  

Keywords:  chemical sterilization; ethylene oxide; hospitals; hydrogen peroxide gas plasma; precautionary practices

Year:  2015        PMID: 26594097      PMCID: PMC4651664     

Source DB:  PubMed          Journal:  Zentralsterilisation (Wiesb)        ISSN: 0942-6086


  8 in total

1.  IARC monographs on the evaluation of carcinogenic risks to humans. Volume 97. 1,3-butadiene, ethylene oxide and vinyl halides (vinyl fluoride, vinyl chloride and vinyl bromide).

Authors: 
Journal:  IARC Monogr Eval Carcinog Risks Hum       Date:  2008

2.  Exposure to hydrogen peroxide and eye and nose symptoms among workers in a beverage processing plant.

Authors:  G Mastrangelo; R Zanibellato; E Fadda; J H Lange; L Scoizzato; R Rylander
Journal:  Ann Occup Hyg       Date:  2008-12-24

3.  Guideline for disinfection and sterilization of prion-contaminated medical instruments.

Authors:  William A Rutala; David J Weber
Journal:  Infect Control Hosp Epidemiol       Date:  2010-02       Impact factor: 3.254

4.  Precautionary practices of healthcare workers who disinfect medical and dental devices using high-level disinfectants.

Authors:  Scott A Henn; James M Boiano; Andrea L Steege
Journal:  Infect Control Hosp Epidemiol       Date:  2015-02       Impact factor: 3.254

5.  NIOSH health and safety practices survey of healthcare workers: training and awareness of employer safety procedures.

Authors:  Andrea L Steege; James M Boiano; Marie H Sweeney
Journal:  Am J Ind Med       Date:  2014-02-18       Impact factor: 2.214

6.  Ethylene oxide burns from improperly sterilized mammary implants.

Authors:  L Cárdenas-Camarena
Journal:  Ann Plast Surg       Date:  1998-10       Impact factor: 1.539

7.  Ethylene oxide exposure may increase the risk of spontaneous abortion, preterm birth, and postterm birth.

Authors:  A S Rowland; D D Baird; D L Shore; B Darden; A J Wilcox
Journal:  Epidemiology       Date:  1996-07       Impact factor: 4.822

8.  Re-evaluation of some organic chemicals, hydrazine and hydrogen peroxide.

Authors: 
Journal:  IARC Monogr Eval Carcinog Risks Hum       Date:  1999
  8 in total
  5 in total

1.  Stability and bioactivity of pepCD47 attachment on stainless steel surfaces.

Authors:  Vaishali V Inamdar; Emmett Fitzpatrick; Ivan Alferiev; Chandrasekaran Nagaswami; Lynn A Spruce; Hossein Fazelinia; George Bratinov; Steven H Seeholzer; Robert J Levy; Ilia Fishbein; Stanley J Stachelek
Journal:  Acta Biomater       Date:  2020-01-11       Impact factor: 8.947

2.  Is the Stella™ 5L system an effective cold sterilization technique for needle-based confocal miniprobes?

Authors:  Cai-Xia Wang; Yuan-Yuan Chen; Feng Yang; Fan Yang; Jin-Tao Guo; Siyu Sun; Ling Fan
Journal:  Endosc Ultrasound       Date:  2017 May-Jun       Impact factor: 5.628

Review 3.  Use of ethylene-oxide gas sterilisation to terminate multidrug-resistant bacterial outbreaks linked to duodenoscopes.

Authors:  Lawrence F Muscarella
Journal:  BMJ Open Gastroenterol       Date:  2019-08-05

4.  Endophthalmitis after pars plana vitrectomy with reused single-use devices: a 13-year retrospective study.

Authors:  Sukhum Silpa-Archa; Kwanchanoke Kumsiang; Janine M Preble
Journal:  Int J Retina Vitreous       Date:  2021-01-06

Review 5.  Decontamination of respirators amid shortages due to SARS-CoV-2.

Authors:  Ravinder Thaper; Brendan Fagen; Jonghwa Oh
Journal:  Photochem Photobiol Sci       Date:  2021-06-12       Impact factor: 4.328

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.