| Literature DB >> 26592650 |
Pinkesh G Sutariya1, Alok Pandya2, Anand Lodha3, Shobhana K Menon4.
Abstract
A new, simple, ultra-sensitive and selective approach has been reported for the "on spot" colorimetric detection of creatinine based on calix[4]arene functionalized gold nanoparticles (AuNPs) with excellent discrimination in the presence of other biomolecules. The lower detection limit of the method is 2.16nM. The gold nanoparticles and p-tert-butylcalix[4]arene were synthesized by microwave assisted method. Specifically, in our study, we used dynamic light scattering (DLS) which is a powerful method for the determination of small changes in particle size, improved selectivity and sensitivity of the creatinine detection system over colorimetric method. The nanoassembly is characterized by Transmission electron microscopy (TEM), DLS, UV-vis and ESI-MS spectroscopy, which demonstrates the binding affinity due its ability of hydrogen bonding and electrostatic interaction between -NH group of creatinine and pSDSC4. It exhibits fast response time (<60s) to creatinine and has long shelf-life (>5 weeks). The developed pSDSC4-AuNPs based creatinine biosensor will be established as simple, reliable and accurate tool for the determination of creatinine in human urine samples.Entities:
Keywords: Biosensor; Calix[4]arene; Creatinine; DLS selectivity; Gold nanoparticles
Mesh:
Substances:
Year: 2015 PMID: 26592650 DOI: 10.1016/j.talanta.2015.10.029
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057