PURPOSE: To describe the successful endovascular treatment of 2 patients with type IIIb endoleak using the Nellix EndoVascular Aneurysm Sealing (EVAS) System. CASE REPORT: Two men aged 75 and 83 years presented with type IIIb endoleak several years (8 and 3, respectively) after initial endovascular aneurysm repair. Dual Nellix endoprosthesis were deployed in the stent-grafts, and the endobags were filled with polymer to seal the defect and eliminate the endoleak. The perioperative periods were uneventful. At up to 6-month follow-up, no endoleaks were detected, aneurysm diameters were unchanged, and the endografts were patent. One patient died 7 months after revision due to a metastatic malignancy.
CONCLUSION: The Nellix EVAS System may prove useful for the treatment of type IIIb endoleak.
© The Author(s) 2015.

Entities:  

Keywords:  abdominal aortic aneurysm; device failure; endoleak; endovascular aneurysm repair; endovascular aneurysm sealing; graft erosion; reintervention; stent-graft

Mesh:

Substances:

Year:  2015        PMID: 26590110     DOI: 10.1177/1526602815617922

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  1 in total

1.  An International, Multicenter Retrospective Observational Study to Assess Technical Success and Clinical Outcomes of Patients Treated with an Endovascular Aneurysm Sealing Device for Type III Endoleak.

Authors:  Aleksandra C Zoethout; Shirley Ketting; Clark J Zeebregts; Dimitri Apostolou; Barend M E Mees; Patrick Berg; Hazem El Beyrouti; Jean-Paul P M De Vries; Francesco Torella; Mattia Migliari; Roberto Silingardi; Michel M P J Reijnen
Journal:  J Endovasc Ther       Date:  2021-08-03       Impact factor: 3.487

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.