Retrieval analysis of titanium nitride (TiN) coated prosthetic femoral heads articulating with polyethylene.

Łukasz Łapaj1, Justyna Wendland2, Jacek Markuszewski3, Adrian Mróz4, Tomasz Wiśniewski4.   

Abstract

Data regarding in vivo performance of titanium nitride (TiN) coated prosthetic femoral heads is scarce, and available studies of older generations of implants demonstrated coating wear in vivo. That is why we conducted a retrieval analysis of 11 femoral heads (articulating in vivo for 1-56 months) with TiN film formed using physical vapor deposition (PVD), to verify if coating failure is a problem in contemporary implants. Retrieved implants were examined using scanning electron microscope, coating roughness was evaluated with a contact profilometer and adhesion was tested using a Rockwell HRC test according to VDI 3824 guideline. Although no gross failure of the TiN coating was observed in our retrievals, all implants had defects typical for PVD coatings, such as pinholes, small titanium droplets and blisters with delaminated coating. In some heads the coating was contaminated with small niobium (Nb) droplets uniformly scattered on the entire surface of the film. Presence of Nb contamination was associated with an increased number and area of other types of defects and poorer coating adhesion. In one component, subjected to multiple dislocations we found severe delamination and cracking of the coating, increased roughness and the presence of third bodies. Our results indicate, that although wear of the coating is lower than seen in older generations of implants, inconsistent quality of the TiN film among different implants indicates the need for strict monitoring of the manufacturing process.
Copyright © 2015 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Coating wear; Orthopedic implant coating; Titanium nitride; Total hip arthroplasty

Mesh:

Substances:

Year:  2015        PMID: 26584076     DOI: 10.1016/j.jmbbm.2015.10.012

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  5 in total

Review 1.  New alternate bearing surfaces in total hip arthroplasty: A review of the current literature.

Authors:  Preston W Grieco; Scott Pascal; Jared M Newman; Neil V Shah; Sarah G Stroud; Neil P Sheth; Aditya V Maheshwari
Journal:  J Clin Orthop Trauma       Date:  2017-10-27

2.  Wear performance of inverted non-conforming bearings in anatomic total shoulder arthroplasty.

Authors:  Lorenzo Banci; Alessio Meoli; Martin Hintner; Hans Rudolf Bloch
Journal:  Shoulder Elbow       Date:  2019-02-07

Review 3.  Updates in biomaterials of bearing surfaces in total hip arthroplasty.

Authors:  Ahmed A Khalifa; Hatem M Bakr
Journal:  Arthroplasty       Date:  2021-11-01

4.  Experimental Investigation of Material Transfer on Bearings for Total Hip Arthroplasty-A Retrieval Study on Ceramic and Metallic Femoral Heads.

Authors:  Jessica Hembus; Lisa Rößler; Armin Springer; Marcus Frank; Annett Klinder; Rainer Bader; Carmen Zietz; Andreas Enz
Journal:  J Clin Med       Date:  2022-07-07       Impact factor: 4.964

5.  A Comparative Study of Friction and Wear Processes of Model Metallic Biomaterials Including Registration of Friction-Induced Temperature Response of a Tribological Pair.

Authors:  Magdalena Łępicka; Artur Ciszewski; Karol Golak; Małgorzata Grądzka-Dahlke
Journal:  Materials (Basel)       Date:  2019-12-11       Impact factor: 3.623

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.