Isotopic Scintigraphy Coupled With Computed Tomography for the Investigation of Intrathecal Baclofen Device Malfunction.

Flavie Frémondière1, Franck Lacoeuille2, Avigaelle Sher2, Olivier Couturier2, Philippe Menei3, Isabelle Richard4, Mickaël Dinomais4.   

Abstract

OBJECTIVE: To assess the potential use of indium-111 diethylenetriamine pentaacetic acid ((111)In-DTPA) scintigraphy coupled with computed tomography (CT) for the investigation of intrathecal baclofen (ITB) device malfunction.
DESIGN: Retrospective study of a case series of patients.
SETTING: Neurosurgical and physical and rehabilitation medicine departments. PARTICIPANTS: Patients (N=7) with reduced ITB effectiveness in whom prior conventional radiographs were inconclusive. INTERVENTION: Nine (111)In-DTPA scintigraphic studies and 8 CT scans. Planar acquisitions were followed by tomoscintigraphy combined with CT. MAIN OUTCOME MEASURE: Progression of the radiotracer in the pump, catheters, and in the subarachnoid space.
RESULTS: In 7 cases, scintigraphy coupled with CT showed leakage behind the pump, lack of activity outside the pump reservoir, abrupt interruption of activity in the catheter, or abnormal distribution of the radiotracer, thus demonstrating that the drug did not reach its target. Surgical revision confirmed these findings in 5 cases. In 1 case, combined imagery ruled out device dysfunction. In the remaining case, only planar acquisitions were performed, showing correct diffusion of the radiotracer.
CONCLUSIONS: The combination of scintigraphy and CT provides simultaneous functional and anatomic imagery of the device. The slow infusion of the radioisotope mimics the diffusion of baclofen, and this could be a useful method to explore intrathecal device malfunction. Further studies are required to compare scintigraphy coupled with CT, to radiopaque injection followed by fluoroscopy or CT.
Copyright © 2016 American Congress of Rehabilitation Medicine. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Baclofen; Cerebral palsy; Equipment failure; Indium; Infusion pumps, implantable; Multiple sclerosis; Muscle spasticity; Radionuclide imaging; Rehabilitation; Spinal cord injuries

Mesh:

Substances:

Year:  2015        PMID: 26582550     DOI: 10.1016/j.apmr.2015.10.101

Source DB:  PubMed          Journal:  Arch Phys Med Rehabil        ISSN: 0003-9993            Impact factor:   3.966


  3 in total

1.  Plain radiography in patients treated with intrathecal drug delivery using an implantable pump device.

Authors:  Elmar M Delhaas; Biswadjiet S Harhangi; Sander P G Frankema; Frank J P M Huygen; Aad van der Lugt
Journal:  Insights Imaging       Date:  2017-08-24

2.  Catheter Access Port (Computed Tomography) Myelography in Intrathecal Drug Delivery Troubleshooting: A Case Series of 70 Procedures.

Authors:  Elmar M Delhaas; Biswadjiet S Harhangi; Sander P G Frankema; Frank J P M Huygen; Aad van der Lugt
Journal:  Neuromodulation       Date:  2020-04-08

3.  Isotopic Scintigraphy in Intrathecal Drug Delivery Failure: A Single-Institution Case Series.

Authors:  Elmar M Delhaas; Daniëlle M E van Assema; Alida C Fröberg; Ben G J C Zwezerijnen; Biswadjiet S Harhangi; Sander P G Frankema; Frank J P M Huygen; Aad van der Lugt
Journal:  Neuromodulation       Date:  2020-09-18