Literature DB >> 26573267

[Randomised study of the relationship between the use of CPRmeter® device and the quality of chest compressions in a simulated cardiopulmonary resuscitation].

J A Calvo-Buey1, D Calvo-Marcos2, R M Marcos-Camina3.   

Abstract

OBJECTIVE: To determine whether the use of CPRmeter(®) during the resuscitation manoeuvres, is related to a higher quality of external cardiac massage, as recommended by the International Liaison Committee on Resuscitation (ILCOR). To compare the quality obtained without the use or this, and whether there are differences related to anthropometric, demographic, professional and/or occupational factors.
METHOD: Experimental, open trial performed with life support simulators in a stratified random sample of 88 health workers randomly distributed between groups A (without indications of the device) and B (with them). The homogeneity of their confounding variables was compared, as well as the compressions depth and compressions rate, the proportion of completed release, and distribution of the quality massage variable (according to criteria ILCOR) between the groups. The qualitative variables were analysed with the chi-square test, and quantitative variables with the Student t-test or Mann-Whitney U-test and the association between the variable quality massage variable, and use of the device with the odds ratio.
RESULTS: Group A: mean depth 42.1mm (standard deviation 10.1), mean rate 121.3/min (21.6), percentage of complete release 71.2% (36.9). Group B: 51.2mm (5.9) 111.9/min (6.4), 92.9% (10.1) respectively. Odds ratio for quality massage regarding the use of the device was 5.170 (95% CI; 2.060-12.977).
CONCLUSIONS: The use of CPRmeter(®) device in simulated resuscitations is related to a higher quality of cardiac massage, improving the approach to the ILCOR recommendations, regardless of the characteristics of the participants. They were 83.8% more likely to achieve a quality massage using the device than without it.
Copyright © 2015 Elsevier España, S.L.U. y SEEIUC. All rights reserved.

Entities:  

Keywords:  Cardiac arrest; Cardiopulmonary resuscitation; Devices; Dispositivos; Heart massage; Masaje cardiaco; Paro cardiaco; Resucitación cardiopulmonar; Simulación; Simulation

Mesh:

Year:  2015        PMID: 26573267     DOI: 10.1016/j.enfi.2015.07.006

Source DB:  PubMed          Journal:  Enferm Intensiva        ISSN: 1130-2399


  2 in total

1.  A comparison of cardiopulmonary resuscitation with standard manual compressions versus compressions with real-time audiovisual feedback: A randomized controlled pilot study.

Authors:  Amir Vahedian-Azimi; Farshid Rahimibashar; Andrew C Miller
Journal:  Int J Crit Illn Inj Sci       Date:  2020-03-06

Review 2.  Real-time audio-visual feedback with handheld nonautomated external defibrillator devices during cardiopulmonary resuscitation for in-hospital cardiac arrest: A meta-analysis.

Authors:  Andrew C Miller; Kiyoshi Scissum; Lorena McConnell; Nathaniel East; Amir Vahedian-Azimi; Kerry A Sewell; Shahriar Zehtabchi
Journal:  Int J Crit Illn Inj Sci       Date:  2020-09-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.