PURPOSE: To test the safety, efficacy, and diagnostic accuracy of automated carbon dioxide (CO2) angiography (ACDA) for the evaluation of diabetic patients with critical limb ischemia (CLI) and baseline renal insufficiency and compare ACDA with iodinated contrast medium (ICM) during endovascular treatment.
METHODS: From November 2014 to January 2015, 36 consecutive diabetic patients (mean age 74.8±5.8 years; 27 men) with stage ≥3 chronic kidney disease (CKD ≥3) and CLI underwent lower limb angiography with both CO2 and ICM followed by balloon angioplasty in a prospective single-center study. The primary outcome measure was the safety and efficacy of ACDA as the exclusive agent to guide angioplasty in this cohort. The secondary outcomes were the safety and diagnostic accuracy of ACDA injection as compared with ICM digital subtraction angiography (DSA) for invasive evaluation of these patients.
RESULTS: ACDA safely and effectively guided angioplasty in all patients without complications. Transcutaneous oxygen pressure improved from 11.8±6.3 to 58.4±7.6 mm Hg (p<0.001). There were no complications related to ACDA during diagnostic imaging and no significant changes in the estimated glomerular filtration rate from baseline to 24 hours (44.7±13.3 vs 47.0±0.8 mL/min/1.73 m(2); nonsignificant). The diagnostic accuracy of CO2 was 89.8% (sensitivity 92.3%; specificity 75%; positive predictive value 95.5%; negative predictive value 63.1%). There was no statistically significant difference in the qualitative diagnostic accuracy between the media (p=0.197).
CONCLUSION: ACDA is an accurate, safe, and effective technique that can be utilized to guide endovascular interventions in diabetics with CLI and baseline CKD ≥3. Larger multicenter randomized studies are needed to validate these results.
© The Author(s) 2015.

Entities:  

Keywords:  angioplasty; automated carbon dioxide angiography; chronic kidney disease; contrast media; critical limb ischemia; diabetic foot; endovascular interventions; iodinated contrast; pedal arteries; transcutaneous oxygen pressure

Mesh:

Substances:

Year:  2015        PMID: 26567126     DOI: 10.1177/1526602815616924

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  4 in total

1.  A Portable Optical Recording Device Simulating CO2 Angiography for Training Purposes.

Authors:  Ivan Corazza; Lorenzo Casadei; Edoardo Pirazzini; Luca Neri; Romano Zannoli
Journal:  J Med Syst       Date:  2017-06-08       Impact factor: 4.460

2.  Efficacy and safety of endovascular therapy by diluted contrast digital subtraction angiography in patients with chronic kidney disease.

Authors:  Naoki Hayakawa; Satoshi Kodera; Noriyoshi Ohki; Junji Kanda
Journal:  Heart Vessels       Date:  2019-04-15       Impact factor: 2.037

Review 3.  Innovations in the Endovascular Management of Critical Limb Ischemia: Retrograde Tibiopedal Access and Advanced Percutaneous Techniques.

Authors:  Jihad A Mustapha; Larry J Diaz-Sandoval; Fadi Saab
Journal:  Curr Cardiol Rep       Date:  2017-08       Impact factor: 2.931

4.  Feasibility and safety of automated CO2 angiography in peripheral arterial interventions.

Authors:  Rohit Philip Thomas; Simon Viniol; Alexander Marc König; Irene Portig; Zaher Swaid; Andreas H Mahnken
Journal:  Medicine (Baltimore)       Date:  2021-01-15       Impact factor: 1.817

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.