Measurement of Femoral Vein Diameter by Ultrasound to Estimate Central Venous Pressure.

Roy J Cho1, David R Williams1, James W Leatherman1.   

Abstract

RATIONALE: Central venous pressure (CVP) can be estimated by ultrasound of the inferior vena cava (IVC), but imaging the IVC is sometimes challenging. The femoral vein is easily imaged by ultrasound and might therefore provide an alternate target for estimating CVP.
OBJECTIVES: To assess femoral vein diameter (FVD) measured by ultrasound imaging for estimating CVP.
METHODS: We prospectively measured CVP and FVD in 97 patients. Receiver operating characteristic curves were used to assess the ability of FVD to predict specific CVP values: less than 10 mm Hg, less than 8 mm Hg (low CVP), and greater than 12 mm Hg (high CVP). Interobserver variability of FVD measurement was assessed in 20 patients.
MEASUREMENTS AND MAIN RESULTS: There was moderate correlation between FVD and CVP (r = 0.66, P < 0.001). FVD less than or equal to 0.8 cm was the best predictor of CVP < 10 mm Hg, with an area under the curve (AUC) of 0.894 and a 95% confidence interval (CI) of 0.82 to 0.97. FVD less than or equal to 0.7 cm performed best for predicting low CVP (AUC = 0.97; 95% CI, 0.94-0.99) and FVD greater than or equal to 1.0 cm for high CVP (AUC = 0.80; 95% CI, 0.72-0.89). However, FVD greater than or equal to 1.2 cm had the greatest specificity (94%) for high CVP. Interobserver variability in FVD measurements was 8.3 ± 7.2%.
CONCLUSIONS: The results of this exploratory study suggest that the accuracy of FVD measured by ultrasound imaging for estimating CVP is comparable to that which has been reported for ultrasound measurement of IVC diameter. FVD may provide an alternative approach when the IVC is difficult to image. Additional studies on other cohorts of patients are warranted to validate our proposed FVD cutoff values for predicting low and high CVP.

Entities:  

Keywords:  central venous pressure; critical care; ultrasound

Mesh:

Year:  2016        PMID: 26561731     DOI: 10.1513/AnnalsATS.201506-337BC

Source DB:  PubMed          Journal:  Ann Am Thorac Soc        ISSN: 2325-6621


  3 in total

1.  Predicting Central Venous Pressure by Measuring Femoral Venous Diameter Using Ultrasonography.

Authors:  Akram Malik; Aftab Akhtar; Shoab Saadat; Salman Mansoor
Journal:  Cureus       Date:  2016-11-23

2.  Measuring the ratio of femoral vein diameter to femoral artery diameter by ultrasound to estimate volume status.

Authors:  Zhihang Ma; Jiaxin Gai; Yinghan Sun; Yunpeng Bai; Hongyi Cai; Lei Wu; Lixiu Sun; Junyan Liu; Li Xue; Bingchen Liu
Journal:  BMC Cardiovasc Disord       Date:  2021-10-20       Impact factor: 2.298

3.  An Inexpensive Cardiovascular Flow Simulator for Cardiac Catheterization Procedure Using a Pulmonary Artery Catheter.

Authors:  Annika Johnson; Grace Cupp; Nicholas Armour; Kyle Warren; Christopher Stone; Davin Lee; Nicholas Gilbert; Chris Hammond; John Moore; Youngbok Abraham Kang
Journal:  Front Med Technol       Date:  2021-10-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.