Literature DB >> 26558277

Auditing of Monitoring and Respiratory Support Equipment in a Level III-C Neonatal Intensive Care Unit.

Elena Bergon-Sendin1, Carmen Perez-Grande1, David Lora-Pablos1, Javier De la Cruz Bertolo1, María Teresa Moral-Pumarega1, Gerardo Bustos-Lozano1, Carmen Rosa Pallas-Alonso1.   

Abstract

BACKGROUND: Random safety audits (RSAs) are a safety tool but have not been widely used in hospitals.
OBJECTIVES: To determine the frequency of proper use of equipment safety mechanisms in relation to monitoring and mechanical ventilation by performing RSAs. The study also determined whether factors related to the patient, time period, or characteristics of the area of admission influenced how the device safety systems were used.
METHODS: A prospective observational study was conducted in a level III-C Neonatal Intensive Care Unit (NICU) during 2012. 87 days were randomly selected. Appropriate overall use was defined when all evaluated variables were correctly programmed in the audited device.
RESULTS: A total of 383 monitor and ventilator audits were performed. The Kappa coefficient of interobserver agreement was 0.93. The rate of appropriate overall use of the monitors and respiratory support equipment was 33.68%. Significant differences were found with improved usage during weekends, OR 1.85 (1.12-3.06, p = 0.01), and during the late shift (3 pm to 10 pm), OR 1.59 (1.03-2.4, p = 0.03).
CONCLUSIONS: Equipment safety systems of monitors and ventilators are not properly used. To improve patient safety, we should identify which alarms are really needed and where the difficulties lie for the correct alarm programming.

Entities:  

Mesh:

Year:  2015        PMID: 26558277      PMCID: PMC4628988          DOI: 10.1155/2015/719497

Source DB:  PubMed          Journal:  Biomed Res Int            Impact factor:   3.411


1. Introduction

Random safety audits (RSAs) are a safety tool widely used in industry due to their ability to identify errors and potentially hazardous situations, thereby enabling prevention of adverse events. RSAs consist of continuously monitoring high-risk procedures in order to identify and address error-prone points in the system, difficult to detect with other methods, before they cause patient harm [1]. The application of this method in a hospital setting could be extremely valuable as it evaluates real-time clinical practice and provides immediate feedback to the staff in the unit [2]. Moreover, this method, which is applied by clinical staff, only requires simple training and involves a low cost of implementation. Despite all these advantages, RSAs have rarely been applied in the hospital setting and there is very little published data on their use as a tool for safety and quality control [3, 4]. Neonatal Intensive Care Units (NICU) have a large variety of different technological devices, most of which have built-in safety systems that help ensure appropriate use. In the field of neonatology, much attention has been paid to other types of error but less to the potential for technology-related errors, despite the fact that good patient outcomes largely depend on how the technology is used. Consequently, we decided to implement RSAs as a safety tool to analyse the use of technology in our NICU. The primary aim of this study was to determine the frequency of proper use of equipment safety mechanisms and other measures involved in patient safety in relation to monitoring and invasive and noninvasive mechanical ventilation by performing RSAs. The study also determined whether factors related to the patient, time period, or characteristics of the area of admission influenced how the device safety systems were used.

2. Materials and Methods

A prospective observational study was conducted from January 1 to December 31, 2012, in a level III-C NICU with 1000 admissions annually. 120 patients with birth weight less than 1500 g per year were admitted to the NICU with a survival rate of 90%, bronchopulmonary dysplasia (BPD) rate at 36 weeks of 16.4%, and a rate for retinopathy of prematurity (ROP) grade 3 or higher of 1.4%. Our NICU is divided into three areas for critical care: a large one with 10 beds (NICU-A) and two small ones (NICU-B and NICU-C), one with 4 and the other with 5 beds, and two additional areas with 24 medium care cots. Physicians and nurses were surveyed according to the modified Delphi technique [5] on the technological devices and procedures in which the recommendations for use were apparently often not met, ensuring that the most relevant equipment and procedures were included. A total of 23 technological devices and procedures were selected (shown in Table 1) and 23 cards were produced, each containing the variables to be evaluated for each device or procedure.
Table 1

All the technological devices and procedures audited in the research. In bold font are the technological devices included in the present study.

Technological devicesProcedures

Masimo RADICAL-7 pulse oximeter Safety
Philips Intellivue MP70 multimeasurement monitor Hand hygiene
Dräger Babylog 8000plus ventilator Prescription of medication
SensorMedics Carefusionhigh-frequency oscillatory ventilator Administration of medication
Viasys Infant Flow system CPAP Administration of enteral nutrition
Dräger Caleo and Ohmeda Giraffe incubatorsAdministration of fluid therapy
Alaris medication pumps Developmental centered care
Alaris fluid therapy pumps Identification and treatment of pain
Alaris enteral nutrition pumps Patient identification
Phototherapy lampsWritten information for parents
Ikaria Nitric Oxide equipment Handling of the lines
Inspiration Healthcare Tecotherm hypothermia equipment
During the study period, 2 days a week were randomly selected and included week days, weekends, and holidays. The shifts, early, from 8 am to 3 pm, or late, from 3 pm to 10 pm, were also randomly selected. Each study day, 2 cards were randomly selected for audit. On one day, 2 kinds of technological devices, 2 kinds of procedures, or 1 kind of device and 1 kind procedure would be audited. As it was used less frequently, the high-frequency oscillatory ventilator was audited every other day whenever it was in use. Each day of the research, a researcher (EBS or MCPG) identified selected devices or procedures in use by patients which had to be audited in order to check all study variables. The NICU staff did not know the purpose of the audit but if an error was detected which might involve a potential danger to the patient, the caregivers were immediately informed. The degree of reproducibility of the 2 investigators was analysed by simultaneous rounds of audits. In this study, the results presented are for the monitoring and respiratory support equipment, so only the audit procedure for this specific equipment is described.

2.1. Monitoring Equipment and Equipment for Invasive and Noninvasive Mechanical Ventilation

Information about Masimo RADICAL-7 pulse oximeters and Philips IntelliVue MP70 multimeasurement monitors was collected (see Table 2). Consideration was given as to whether the monitored patients had some type of assisted ventilation and supplemental oxygen. The outcome variable “appropriate overall use” was defined to describe when all the variables were correctly programmed in the audited monitor. Information about Dräger Babylog 8000plus conventional mechanical ventilator, Carefusion SensorMedics high-frequency oscillatory ventilator, and Viasys Infant Flow system continuous positive airway pressure (CPAP) was collected (see Table 2). The outcome variable “appropriate overall use” was defined to describe when all the variables were correctly programmed in the audited ventilator. A critical incident was defined as any situation detected concerning the devices reviewed that could put the patient in immediate risk.
Table 2

Definitions of appropriate programming of the variables assessed for the technological devices. ECG: electrocardiogram; PIP: peak inflating pressure; MAP: mean airway pressure; CPAP: continuous positive airway pressure; PEEP: positive end-expiratory pressure.

Appropriate programming
Pulse oximeterSpO2 alarmsLower limit 85% and upper limit 95% (as per protocol) or as instructed by a doctor
Heart rate alarmsLower limit 95 and upper limit 195 bpm or as instructed by a doctor
Alarm volume50–70 dB
Sensor statusClearly defined curves without artefacts

Multimeasurement monitorSpO2 alarmsLower limit 85% and upper limit 95% (as per protocol) or as instructed by a doctor
Heart rate alarmsLower limit 95 and upper limit 195 bpm or as instructed by a doctor [6]
Respiratory rate alarmsLower limit 30 and upper limit 70 rpm or as instructed by a doctor [6]
Blood pressure alarms±20% of the normal value for the patient's gestational age [6]
Alarm volume50–70 dB
Sensor status and ECG padsClearly defined curves without artefacts

Conventional mechanical ventilatorVolume/minute alarm±20% of the normal volume/minute for a newborn [7, 8]
Apnoea alarm20 seconds
Alarm volume50–70 dB
Peak limit in volume guarantee modeAt least 5 to 10 cm H2O above the working PIP [9]
Respirator flow6-7 lpm for number 2.5 endotracheal tubes
8–10 lpm for number 3 and number 3.5 endotracheal tubes
Position of the flow sensorSlightly above the level of the patient
HeaterWith water and temperature set at 37°C
Change of tubingAs per unit policy (<7 days)

High-frequency oscillatory ventilatorMAP alarms±3 points of the programmed MAP
Position of tubingRunning straight and at the height of the patient
HeaterWith water and temperature set at 37°C

CPAPPEEPAs prescribed
HeaterWith water and temperature set at 39°C
Change of tubingAs per unit policy (<7 days)

2.2. Other Variables

Information was collected on the patient, the time, and characteristics of the NICU area where the patient was admitted in order to evaluate whether or not they influenced the use of the safety systems of these devices: birth weight; gestational age; sex; occupancy of the NICU at the time of the audit; and location of the patient.

2.3. Work Organization in the NICU

In the NICU, the ratio patient/nurse is 2.1 during all the shifts. The nurses are registered nurses with university education and neonatal experience. Paediatric nurse specialists were introduced 3 years ago in Spain. In relation to the doctors, 7 neonatologist and 7 paediatrics residents are in charge of the 19 intensive care beds and 24 medium care beds. During the duty (from 3 pm to 8 am) 2 neonatologist consultants and 1 resident are in charge of the neonatal unit including attention to delivery room. The nurses are responsible for programming monitor and pulse oximeter alarms following the unit's protocols or medical instructions, for controlling the water and heater temperatures, and for changing respirator and CPAP tubing. The respirator and CPAP parameters and alarms are programmed by doctors.

2.4. Ethical Issues

This study involved quality strategies for improving patient security and thus did not require institutional review board approval. The study consent was obtained from the head doctor and the head nurses of the unit.

2.5. Analysis Plan

Continuous variables are presented as mean ± SD and categorical variables as absolute and relative frequencies. The reproducibility of the observations made by the two study investigators was estimated with the kappa coefficient. The statistical significance of the comparison of proportions was determined using chi-square or Fisher's exact test from contingency tables. Comparisons of the distributions of ordinal and continuous measurements were made using the Wilcoxon-Mann-Whitney test or Student's t-test, as appropriate. Multivariable logistic regression analysis was used to estimate the strength of association between appropriate overall use and several covariables such as gestational age, birth weight, sex, location in the NICU, normal working day or nonworking day, shift, month, and occupancy. Results are presented as odds ratios and 95% confidence intervals (CI).

3. Results

During the study period, a total of 383 assessments were made of the monitoring and respiratory support equipment (see Figure 1). The overall interobserver kappa coefficient between the 2 investigators carrying out the audits was 0.93. The rate of appropriate use of the equipment safety systems and other aspects of the technology as a whole relating to patient safety was 33.68% (129/383). Table 3 shows the rates for appropriate use of the items assessed in each device.
Figure 1

Flowchart of the audits focusing on the assessment of equipment and procedures during the study period. This study is part of a broader investigation that included other technological devices which are shown in Table 1.

Table 3

Rates of appropriate use for the variables evaluated in each device and technology as a whole. SpO2: oxygen saturation; VG: volume guarantee ventilation; MAP: mean airway pressure; CPAP: continuous positive airway pressure; PEEP: positive end-expiratory pressure.

Technological devices (N assessments)Appropriate use (N)Appropriate use % (confidence interval)
Pulse oximeters (N 94)
 SpO2 alarms61 64.89 (55.24–74.54)
 Heart rate alarms87 92.55 (87.25–97.86)
 Alarm volume94 100 (96.15–100)
 Sensor status 94 100 (96.15–100)
 Appropriate overall use 58 61.70 (51.88–71.53)
Monitors (N 104)
 SpO2 alarms (N 104)58 55.77 (46.22–65.31)
 Heart rate alarms (N 104)84 80.77 (73.19–88.34)
 Respiratory rate alarms (N 66)12 18.18 (8.87–27.49)
 Blood pressure alarms (N 64)30 46.88 (34.49–59.10)
 Alarm volume (N 104)104 100 (96.51–100)
 State of ECG pads (N 75)68 90.67 (84.08–97.25)
 Appropriate overall use (N 104)22 21.15 (13.30–29.00)
Conventional mechanical ventilator (N 53)
 Volume/minute alarms 18 33.96 (21.21–46.71)
 Apnoea alarm48 90.57 (82.69–98.44)
 Peak pressure limit in VG (N 42)38 90.48 (81.59–99.35)
 Flow appropriate to the endotracheal tube size48 90.57 (82.69–98.44)
 Adequate position of the flow sensor33 62.26 (49.21–75.31)
 Change of tubing26 49.06 (35.59–62.52)
 Alarm volume 53 100 (93.27–100)
 Heater temperature53 100 (93.27–100)
 Water in the heater52 98.11 (94.45–100)
 Appropriate overall use 611.32 (2.79–19.85)
High-frequency oscillatory ventilator (N 64)
 MAP alarms33 51.56 (39.32–63.81)
 Position of the tubing 46 71.88 (60.86–82.89)
 Heater temperature63 98.44 (95.40–100)
 Water in the heater61 95.31 (90.13–100)
 Appropriate overall use 24 37.50 (22.64–49.36)
CPAP (N 68)
 PEEP 63 92.65 (86.44–98.85)
 Change of tubing33 48.53 (36.65–60.41)
 Heater temperature 42 61.76 (50.21–73.32)
 Water in the heater68 100 (94.72–100)
 Appropriate overall use 1927.94 (17.28–38.61)
Technology as a whole (N 383)12933.68 (28.81–38.54)
The variables with the worst compliance rates were in the pulse oximeters, the programming of the oxygen saturation (SpO2) alarms at 64.89% (61/94); in the monitors, the respiratory rate alarm at 18.18% (12/104); in the conventional mechanical ventilation respirators, the volume/minute alarm at 33.96% (18/53); in the high-frequency oscillatory ventilator, the MAP alarm at 51.56% (33/64); and in CPAP, the change of tubing at 48.53% (33/68). Table 4 shows the results for appropriate overall use of the different pieces of equipment in relation to patient characteristics and location. Table 5 shows the results in relation to time characteristics. Significant differences were found in the multivariate analysis, with improved overall usage during weekends and holidays, OR 1.85 (1.12–3.06, p = 0.01), and on the late shift, OR 1.59 (1.03–2.4, p = 0.03). The occupancy rate when appropriate use was found was 95.56% and 93.02% when inadequate use was detected (p = 0.002). There were no significant differences in any of the other variables studied.
Table 4

Rates for appropriate overall use and variable with lower rates of appropriate programming of the different pieces of equipment in relation to patient characteristics and location in the Neonatal Intensive Care Unit. NICU: Neonatal Intensive Care Unit; SpO2: oxygen saturation; MAP: mean airway pressure; CPAP: continuous positive airway pressure.

Technological resources (N assessments) Birth weight (g) (N assessments) Gestational age (weeks)(N assessments)Sex (N assessments)Location (N assessments)
<1500≥1500 p <32 ≥32 p MaleFemale p NICU-ANICU-B + C p
Pulse oximeter (N 94)
 % appropriate overall use (N)66 (33) 55.81 (24) 0.31 64.29 (36) 57.89 (22) 0.53 60.47 (26) 62.75 (32) 0.82 57.78 (26) 65.31 (32) 0.45
Multimeasurement monitor (N 104)
 % appropriate overall use (N)24 (12) 18.52 (10) 0.49 26.32 (15) 14.89 (7)0.15 26.53 (13) 16.36 (9)0.20 18.33 (11) 25 (11) 0.41
Conventional mechanical ventilator (N 53)
 % appropriate overall use (N)0.09 (2)13.79 (4)0.60 10 (3)13.64 (3)0.68 10 (3)13.04 (3)0.72 13.89 (5)5.88 (1)0.39
High-frequency oscillatory ventilator (N 64)
 % appropriate overall use (N)32.5 (13) 45.83 (11) 0.28 31.82 (14) 50 (10) 0.16 32.43 (12) 44.44 (12) 0.32 54.17 (13) 27.5 (11) 0.03
CPAP (N 68)
 % appropriate overall use (N)28.89 (13) 27.27 (6)0.89 29.63 (16) 21.43 (3)0.54 22.86 (8)33.33 (11) 0.33 22.73 (5)30.43 (14) 0.50
Technology as a whole (383)
 % appropriate overall use (N)35.27 (73) 31.98 (55) 0.50 34.85 (84) 31.91 (45) 0.55 31.96 (62) 35.45 (67) 0.46 32.09 (60) 35.20 (69) 0.51
Table 5

Rates for appropriate overall use and variable with lower rates of appropriate programming of the different pieces of equipment in relation to time characteristics. SpO2: oxygen saturation; MAP: mean airway pressure; Vol/min alarms: volume/minute alarms; CPAP: continuous positive airway pressure.

Technological resources (N assessments)DayShiftQuarter
Normal working dayNonworking day p EarlyLate p 3rd quarterRest of year p
Pulse oximeter (N 94)
 % appropriate overall use (N)64.56 (51) 46.67 (7)0.19 65.12 (28) 58.82 (30) 0.53 76.92 (20) 55.88 (38) 0.06
Multimeasurement monitor (N 104)
 % appropriate overall use (N)16.05 (13) 39.13 (9)0.01 16.22 (12) 33.33 (10) 0.05 19.35 (6)21.92 (16) 0.76
Conventional mechanical ventilator (N 53)
 % appropriate overall use (N)8.51 (4)33.33 (2)0.07 10.34 (3)12.50 (3)0.80 18.75 (3)8.11 (3)0.26
High-frequency oscillatory ventilator (N 64)
 % appropriate overall use (N)25.53 (12) 70.59 (12) 0.001 38.46 (15) 36 (9)0.84 35 (7)38.64 (17) 0.78
CPAP (N 68)
 % appropriate overall use (N)27.91 (12) 28 (7)0.99 22.86 (8)33.33 (11) 0.33 35.71 (5)25.93 (14) 0.46
Technology as a whole (N 383)
 % appropriate overall use (N)30.98 (92) 43.02 (37) 0.03 30 (66) 38.65 (63) 0.07 38.32 (41) 31.88 (88) 0.23
Critical incidents were found in 9 (2.34%) of the 383 devices reviewed. These were pulse oximeter not working in 2 patients requiring monitoring; alarms not programmed in invasive blood pressure monitoring in 3 unstable patients; and no water in the heater of 4 respirators.

4. Discussion

This prospective observational study shows that only a third of evaluated monitors and respiratory support equipment had all the parameters well programmed. Contrary to what might be expected, weekend and holidays and late shifts, when there are fewer neonatologists in the NICU, were the periods when we found that the evaluated equipment parameters were best programmed. During the study period the NICU occupancy rate was higher than 90%. Patient characteristics and location in the unit did not influence the appropriate use of the monitors and respiratory support equipment. To our knowledge, this is the first published study with data on real-time audits of monitoring and respiratory support equipment in a level III-C NICU. Compared to the large number of published studies on other aspects of neonatal patient safety [10-12], there are surprisingly few literature references reporting errors directly related to the equipment used [3, 13, 14]. For this reason, it has not been possible to compare many of our results with other studies. When we designed the study, we believed that our results were going to be much better than those we finally obtained after the audits. One of the most disturbing findings in relation to the monitoring equipment was the poor protocol compliance for programming SpO2 alarms, results similar to those found by other authors [3, 13]. In most cases, this consisted of not programming the upper SpO2 alarm limit in patients on supplemental oxygen, which exposes the patient to a greater risk of hyperoxia and the widely described adverse effects on lung, retina, and brain [15-17]. There is currently a great deal of concern about what the optimal SpO2 ranges are for premature infants and numerous studies are being conducted in this area [16-19]. However, all the efforts will be of little effect if, on implementing the recommendations in different NICUs, the SpO2 alarm limits are not properly set. Despite the high rate of inadequate programming of SpO2 alarms, our BPD and grade 3 or higher ROP rates are not high. Nevertheless, this does not justify the poor compliance in programming these alarms in view of the other repercussions on the newborns in our unit. One of the most surprising results was finding a higher percentage of appropriate overall use at the weekend and during the late shift. It is not easy to explain this result. At the weekend and during late shifts, there are only two neonatologists on duty. The number of nurses is always the same on all shifts. Programmed procedures are usually not performed. There are, therefore, fewer interruptions, allowing a greater degree of concentration on the part of the staff. The nurses probably work better in the quieter, calmer atmosphere on the weekend. These aspects need to be confirmed since it does not seem logical that the use of technology should be better when there are fewer doctors working. Also remarkable is that appropriate use of the equipment is better when there is higher occupancy. A tendency towards better use would be expected with lower occupancy and not the other way round. In our NICU, the occupancy rate is always very high and although the difference is statistically significant, it may not be clinically relevant. We also wanted to find out whether the use of technology was different in the months of July/August/September compared to the rest of the year, as this is considered a high-risk period for adverse events due to the fact that, in Spain, it is a vacation period during which less-experienced nurses are recruited. For example, our group has reported a higher rate of unplanned extubation over the summer months [20]. However, no differences were found in the use of technology. With regard to adverse events, higher rates have been reported in the most premature patients and those weighing less than 1500 g [21]. In our study, these patient characteristics had no influence on the use of the equipment safety systems or on other safety-related aspects of the technology evaluated. The incidents found in 2.34% of the devices reviewed were considered critical for the serious consequences they could have had for the patient. This rate is not insignificant and NICUs are highly complex units, in which high-risk situations continually occur. In all the above cases, the patient's care provider was immediately informed to resolve the problem. A major problem in the NICUs is the large number of alarms used which produce a high level of noise. There are reports that exposure to high levels of noise in newborns, particularly when premature, can have potentially negative effects on their stability and their future neurodevelopment [22, 23]. Moreover, there are studies which show that 72–99% of the alarms that go off are not actually related to any real patient event [24]. This can lead to a situation where NICU staff suffer from “clinical alarm fatigue,” the term used to describe the phenomenon of medical providers becoming too desensitized due to the constant noise emitted from monitors [25, 26]. It could also lead to staff becoming less rigorous and not programming alarms which they believe often sound without signifying a real clinical problem for the patient. In light of the above, there is an attempt nowadays to design “smart alarms” which manage information in an integrated manner using algorithms, thereby improving the sensitivity and specificity of the alerts [27]. One limitation to our study was that although the staffs were unaware of the objectives of the observation they were, however, aware of being watched and this may have influenced their usual practices. Nevertheless, we do not believe that this aspect influenced the results. When the different quarters of the year were compared, no changes were found in staff behaviour over the course of the study period. Another limitation is that the night shift was not included in the observations. Perhaps our results cannot be generalized because the organization of other units and the degree of professional training may differ between countries and units, but these results may encourage other units to audit the technology they use in an effort to identify areas for improvement. The results of this study have been disseminated to the NICU staff and different strategies have been introduced to improve the appropriate use of equipment. Probably, as in other aspects concerning safety, checklists are useful to evaluate the appropriate settings of monitors and respirators. The companies that design monitoring and respiratory support systems are trying to incorporate an increasing number of alerts to ensure proper use of the equipment and to improve the safety of the newborn. However, our study shows that a large number of these safety systems are not properly used and do not benefit the patient. To improve patient safety, we should seek to identify which alarms are really needed and where the difficulties lie for the correct alarm programming in the monitors and support respiratory equipment.
  26 in total

1.  [Recommendations for respiratory support in the newborn].

Authors: 
Journal:  An Pediatr (Barc)       Date:  2012-05-09       Impact factor: 1.500

2.  Health care failure mode and effect analysis to reduce NICU line-associated bloodstream infections.

Authors:  Celeste J Chandonnet; Prerna S Kahlon; Pratik Rachh; Michele Degrazia; Eileen C Dewitt; Kathleen A Flaherty; Nadine Spigel; Stephanie Packard; Denise Casey; Christine Rachwal; Pankaj B Agrawal
Journal:  Pediatrics       Date:  2013-05-20       Impact factor: 7.124

3.  A practical guide to neonatal volume guarantee ventilation.

Authors:  C Klingenberg; K I Wheeler; P G Davis; C J Morley
Journal:  J Perinatol       Date:  2011-07-14       Impact factor: 2.521

4.  Medication errors and adverse drug events in pediatric inpatients.

Authors:  R Kaushal; D W Bates; C Landrigan; K J McKenna; M D Clapp; F Federico; D A Goldmann
Journal:  JAMA       Date:  2001-04-25       Impact factor: 56.272

5.  Alarm fatigue.

Authors:  Sue Sendelbach
Journal:  Nurs Clin North Am       Date:  2012-07-04       Impact factor: 1.208

6.  Audits in real time for safety in critical care: definition and pilot study.

Authors:  G Sirgo Rodríguez; M Olona Cabases; M C Martin Delgado; F Esteban Reboll; A Pobo Peris; M Bodí Saera
Journal:  Med Intensiva       Date:  2014-02-05       Impact factor: 2.491

7.  Alarm fatigue: a patient safety concern.

Authors:  Sue Sendelbach; Marjorie Funk
Journal:  AACN Adv Crit Care       Date:  2013 Oct-Dec

8.  [Incidence of unplanned extubations in a neonatal intensive care unit. A before and after study].

Authors:  M I Utrera Torres; M T Moral Pumarega; N R García Lara; A Melgar Bonís; M E Frías García; C R Pallás Alonso
Journal:  An Pediatr (Barc)       Date:  2013-10-05       Impact factor: 1.500

9.  Oxygen saturation and outcomes in preterm infants.

Authors:  Ben J Stenson; William O Tarnow-Mordi; Brian A Darlow; John Simes; Edmund Juszczak; Lisa Askie; Malcolm Battin; Ursula Bowler; Roland Broadbent; Pamela Cairns; Peter Graham Davis; Sanjeev Deshpande; Mark Donoghoe; Lex Doyle; Brian W Fleck; Alpana Ghadge; Wendy Hague; Henry L Halliday; Michael Hewson; Andrew King; Adrienne Kirby; Neil Marlow; Michael Meyer; Colin Morley; Karen Simmer; Win Tin; Stephen P Wardle; Peter Brocklehurst
Journal:  N Engl J Med       Date:  2013-05-05       Impact factor: 91.245

10.  Effects of targeting higher vs lower arterial oxygen saturations on death or disability in extremely preterm infants: a randomized clinical trial.

Authors:  Barbara Schmidt; Robin K Whyte; Elizabeth V Asztalos; Diane Moddemann; Christian Poets; Yacov Rabi; Alfonso Solimano; Robin S Roberts
Journal:  JAMA       Date:  2013-05-22       Impact factor: 56.272

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.