| Literature DB >> 26558277 |
Elena Bergon-Sendin1, Carmen Perez-Grande1, David Lora-Pablos1, Javier De la Cruz Bertolo1, María Teresa Moral-Pumarega1, Gerardo Bustos-Lozano1, Carmen Rosa Pallas-Alonso1.
Abstract
BACKGROUND: Random safety audits (RSAs) are a safety tool but have not been widely used in hospitals.Entities:
Mesh:
Year: 2015 PMID: 26558277 PMCID: PMC4628988 DOI: 10.1155/2015/719497
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
All the technological devices and procedures audited in the research. In bold font are the technological devices included in the present study.
| Technological devices | Procedures |
|
| |
|
| Safety |
|
| Hand hygiene |
|
| Prescription of medication |
|
| Administration of medication |
|
| Administration of enteral nutrition |
| Dräger Caleo and Ohmeda Giraffe incubators | Administration of fluid therapy |
| Alaris medication pumps | Developmental centered care |
| Alaris fluid therapy pumps | Identification and treatment of pain |
| Alaris enteral nutrition pumps | Patient identification |
| Phototherapy lamps | Written information for parents |
| Ikaria Nitric Oxide equipment | Handling of the lines |
| Inspiration Healthcare Tecotherm hypothermia equipment | |
Definitions of appropriate programming of the variables assessed for the technological devices. ECG: electrocardiogram; PIP: peak inflating pressure; MAP: mean airway pressure; CPAP: continuous positive airway pressure; PEEP: positive end-expiratory pressure.
| Appropriate programming | ||
|---|---|---|
| Pulse oximeter | SpO2 alarms | Lower limit 85% and upper limit 95% (as per protocol) or as instructed by a doctor |
| Heart rate alarms | Lower limit 95 and upper limit 195 bpm or as instructed by a doctor | |
| Alarm volume | 50–70 dB | |
| Sensor status | Clearly defined curves without artefacts | |
|
| ||
| Multimeasurement monitor | SpO2 alarms | Lower limit 85% and upper limit 95% (as per protocol) or as instructed by a doctor |
| Heart rate alarms | Lower limit 95 and upper limit 195 bpm or as instructed by a doctor [ | |
| Respiratory rate alarms | Lower limit 30 and upper limit 70 rpm or as instructed by a doctor [ | |
| Blood pressure alarms | ±20% of the normal value for the patient's gestational age [ | |
| Alarm volume | 50–70 dB | |
| Sensor status and ECG pads | Clearly defined curves without artefacts | |
|
| ||
| Conventional mechanical ventilator | Volume/minute alarm | ±20% of the normal volume/minute for a newborn [ |
| Apnoea alarm | 20 seconds | |
| Alarm volume | 50–70 dB | |
| Peak limit in volume guarantee mode | At least 5 to 10 cm H2O above the working PIP [ | |
| Respirator flow | 6-7 lpm for number 2.5 endotracheal tubes | |
| 8–10 lpm for number 3 and number 3.5 endotracheal tubes | ||
| Position of the flow sensor | Slightly above the level of the patient | |
| Heater | With water and temperature set at 37°C | |
| Change of tubing | As per unit policy (<7 days) | |
|
| ||
| High-frequency oscillatory ventilator | MAP alarms | ±3 points of the programmed MAP |
| Position of tubing | Running straight and at the height of the patient | |
| Heater | With water and temperature set at 37°C | |
|
| ||
| CPAP | PEEP | As prescribed |
| Heater | With water and temperature set at 39°C | |
| Change of tubing | As per unit policy (<7 days) | |
Figure 1Flowchart of the audits focusing on the assessment of equipment and procedures during the study period. This study is part of a broader investigation that included other technological devices which are shown in Table 1.
Rates of appropriate use for the variables evaluated in each device and technology as a whole. SpO2: oxygen saturation; VG: volume guarantee ventilation; MAP: mean airway pressure; CPAP: continuous positive airway pressure; PEEP: positive end-expiratory pressure.
| Technological devices | Appropriate use ( | Appropriate use % |
|---|---|---|
| Pulse oximeters ( | ||
| SpO2 alarms | 61 | 64.89 (55.24–74.54) |
| Heart rate alarms | 87 | 92.55 (87.25–97.86) |
| Alarm volume | 94 | 100 (96.15–100) |
| Sensor status | 94 | 100 (96.15–100) |
| Appropriate overall use | 58 | 61.70 (51.88–71.53) |
| Monitors ( | ||
| SpO2 alarms ( | 58 | 55.77 (46.22–65.31) |
| Heart rate alarms ( | 84 | 80.77 (73.19–88.34) |
| Respiratory rate alarms ( | 12 | 18.18 (8.87–27.49) |
| Blood pressure alarms ( | 30 | 46.88 (34.49–59.10) |
| Alarm volume ( | 104 | 100 (96.51–100) |
| State of ECG pads ( | 68 | 90.67 (84.08–97.25) |
| Appropriate overall use ( | 22 | 21.15 (13.30–29.00) |
| Conventional mechanical ventilator ( | ||
| Volume/minute alarms | 18 | 33.96 (21.21–46.71) |
| Apnoea alarm | 48 | 90.57 (82.69–98.44) |
| Peak pressure limit in VG ( | 38 | 90.48 (81.59–99.35) |
| Flow appropriate to the endotracheal tube size | 48 | 90.57 (82.69–98.44) |
| Adequate position of the flow sensor | 33 | 62.26 (49.21–75.31) |
| Change of tubing | 26 | 49.06 (35.59–62.52) |
| Alarm volume | 53 | 100 (93.27–100) |
| Heater temperature | 53 | 100 (93.27–100) |
| Water in the heater | 52 | 98.11 (94.45–100) |
| Appropriate overall use | 6 | 11.32 (2.79–19.85) |
| High-frequency oscillatory ventilator ( | ||
| MAP alarms | 33 | 51.56 (39.32–63.81) |
| Position of the tubing | 46 | 71.88 (60.86–82.89) |
| Heater temperature | 63 | 98.44 (95.40–100) |
| Water in the heater | 61 | 95.31 (90.13–100) |
| Appropriate overall use | 24 | 37.50 (22.64–49.36) |
| CPAP ( | ||
| PEEP | 63 | 92.65 (86.44–98.85) |
| Change of tubing | 33 | 48.53 (36.65–60.41) |
| Heater temperature | 42 | 61.76 (50.21–73.32) |
| Water in the heater | 68 | 100 (94.72–100) |
| Appropriate overall use | 19 | 27.94 (17.28–38.61) |
| Technology as a whole ( | 129 | 33.68 (28.81–38.54) |
Rates for appropriate overall use and variable with lower rates of appropriate programming of the different pieces of equipment in relation to patient characteristics and location in the Neonatal Intensive Care Unit. NICU: Neonatal Intensive Care Unit; SpO2: oxygen saturation; MAP: mean airway pressure; CPAP: continuous positive airway pressure.
| Technological resources ( |
Birth weight (g) |
Gestational age (weeks) | Sex ( | Location ( | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| <1500 | ≥1500 |
| <32 | ≥32 |
| Male | Female |
| NICU-A | NICU-B + C |
| |
| Pulse oximeter ( | ||||||||||||
| % appropriate overall use ( | 66 (33) | 55.81 (24) | 0.31 | 64.29 (36) | 57.89 (22) | 0.53 | 60.47 (26) | 62.75 (32) | 0.82 | 57.78 (26) | 65.31 (32) | 0.45 |
| Multimeasurement monitor ( | ||||||||||||
| % appropriate overall use ( | 24 (12) | 18.52 (10) | 0.49 | 26.32 (15) | 14.89 (7) | 0.15 | 26.53 (13) | 16.36 (9) | 0.20 | 18.33 (11) | 25 (11) | 0.41 |
| Conventional mechanical ventilator ( | ||||||||||||
| % appropriate overall use ( | 0.09 (2) | 13.79 (4) | 0.60 | 10 (3) | 13.64 (3) | 0.68 | 10 (3) | 13.04 (3) | 0.72 | 13.89 (5) | 5.88 (1) | 0.39 |
| High-frequency oscillatory ventilator ( | ||||||||||||
| % appropriate overall use ( | 32.5 (13) | 45.83 (11) | 0.28 | 31.82 (14) | 50 (10) | 0.16 | 32.43 (12) | 44.44 (12) | 0.32 | 54.17 (13) | 27.5 (11) | 0.03 |
| CPAP ( | ||||||||||||
| % appropriate overall use ( | 28.89 (13) | 27.27 (6) | 0.89 | 29.63 (16) | 21.43 (3) | 0.54 | 22.86 (8) | 33.33 (11) | 0.33 | 22.73 (5) | 30.43 (14) | 0.50 |
| Technology as a whole (383) | ||||||||||||
| % appropriate overall use ( | 35.27 (73) | 31.98 (55) | 0.50 | 34.85 (84) | 31.91 (45) | 0.55 | 31.96 (62) | 35.45 (67) | 0.46 | 32.09 (60) | 35.20 (69) | 0.51 |
Rates for appropriate overall use and variable with lower rates of appropriate programming of the different pieces of equipment in relation to time characteristics. SpO2: oxygen saturation; MAP: mean airway pressure; Vol/min alarms: volume/minute alarms; CPAP: continuous positive airway pressure.
| Technological resources ( | Day | Shift | Quarter | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Normal working day | Nonworking day |
| Early | Late |
| 3rd quarter | Rest of year |
| |
| Pulse oximeter ( | |||||||||
| % appropriate overall use ( | 64.56 (51) | 46.67 (7) | 0.19 | 65.12 (28) | 58.82 (30) | 0.53 | 76.92 (20) | 55.88 (38) | 0.06 |
| Multimeasurement monitor ( | |||||||||
| % appropriate overall use ( | 16.05 (13) | 39.13 (9) | 0.01 | 16.22 (12) | 33.33 (10) | 0.05 | 19.35 (6) | 21.92 (16) | 0.76 |
| Conventional mechanical ventilator ( | |||||||||
| % appropriate overall use ( | 8.51 (4) | 33.33 (2) | 0.07 | 10.34 (3) | 12.50 (3) | 0.80 | 18.75 (3) | 8.11 (3) | 0.26 |
| High-frequency oscillatory ventilator ( | |||||||||
| % appropriate overall use ( | 25.53 (12) | 70.59 (12) | 0.001 | 38.46 (15) | 36 (9) | 0.84 | 35 (7) | 38.64 (17) | 0.78 |
| CPAP ( | |||||||||
| % appropriate overall use ( | 27.91 (12) | 28 (7) | 0.99 | 22.86 (8) | 33.33 (11) | 0.33 | 35.71 (5) | 25.93 (14) | 0.46 |
| Technology as a whole ( | |||||||||
| % appropriate overall use ( | 30.98 (92) | 43.02 (37) | 0.03 | 30 (66) | 38.65 (63) | 0.07 | 38.32 (41) | 31.88 (88) | 0.23 |