Literature DB >> 26555583

Single-Pass Microkeratome System for Eye Bank DSAEK Tissue Preparation: Is Stromal Bed Thickness Predictable and Reproducible?

Mazen Y Choulakian1, Jennifer Y Li, Samuel Ramos, Mark J Mannis.   

Abstract

PURPOSE: To evaluate the predictability and reproducibility of stromal bed thickness for single-pass donor Descemet stripping automated endothelial keratoplasty (DSAEK) tissue preparation, using the ML7 Microkeratome Donor Cornea System (Med-logics Inc, Athens, TX).
METHODS: In this retrospective chart review of 256 consecutive corneal tissue preparations for DSAEK surgery, from June 2013 to August 2014, tissue thicknesses were divided into 3 groups, depending on surgeon preference: <91 μm (group A), 90 to 120 μm (group B), and 120 to 160 μm (group C). Precut and postcut data were recorded.
RESULTS: Average postcut donor corneal thickness was 114 ± 30 μm (range 60-183 μm), whereas the average in group A was 97 ± 23 μm (range 60-128), in group B was 113 ± 21 μm (range 77-179), and in group C was 134 ± 43 (range 89-183). Average postcut endothelial cell density was very adequate at 3013 ± 250 cells per square millimeter. There were a total of 7 failed procedures from 256 attempts, which represents a rate of 2.7%. This rate decreases to 1.5% when analyzing the last 200 cuts.
CONCLUSIONS: The ML7 Microkeratome Donor Cornea System allows for reliable and reproducible DSAEK tissue preparation. Ultrathin DSAEK tissues can be prepared with a single-pass. Aiming for a graft thickness between 90 and 120 μm seems to be most reliable.

Entities:  

Mesh:

Year:  2016        PMID: 26555583     DOI: 10.1097/ICO.0000000000000675

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  3 in total

1.  Prelamellar Dissection Donor Corneal Thickness Is Associated With Descemet Stripping Automated Endothelial Keratoplasty Operative Complications in the Cornea Preservation Time Study.

Authors:  Kevin W Ross; Christopher G Stoeger; George O D Rosenwasser; Robert C OʼBrien; Loretta B Szczotka-Flynn; Allison R Ayala; Maureen G Maguire; Beth Ann Benetz; Patricia Dahl; Donna C Drury; Steven P Dunn; Sameera M Farazdaghi; Caroline K Hoover; Marian S Macsai; Shahzad I Mian; Michael L Nordlund; Jeffrey G Penta; Mark C Soper; Mark A Terry; David D Verdier; Doyce V Williams; Jonathan H Lass
Journal:  Cornea       Date:  2019-09       Impact factor: 3.152

2.  Prospective evaluation of clinical outcomes between pre-cut corneal grafts prepared using a manual or automated technique: with one-year follow-up.

Authors:  Marc B Muijzer; Chantal M van Luijk; Antoon J van den Bogaerdt; Piet J Kruit; Esther Groeneveld-van Beek; Gerrit R J Melles; Robert P L Wisse
Journal:  Acta Ophthalmol       Date:  2019-02-25       Impact factor: 3.761

3.  Femtosecond laser semi-assisted Descemet stripping endothelial keratoplasty: 2-year outcomes of endothelial cell loss and graft survival.

Authors:  Ting Wang; Peiyan Shi; Fengjie Li; Hua Gao; Suxia Li; Tong Liu; Weiyun Shi
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-08-31       Impact factor: 3.117

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.