Cell Population Data and reflex testing rules of cell analysis in pleural and ascitic fluids using body fluid mode on Sysmex XN-9000.

Sabrina Buoro1, Tommaso Mecca2, Giovanna Azzarà2, Michela Seghezzi2, Paola Dominoni2, Alberto Crippa2, Cosimo Ottomano3, Giuseppe Lippi4.   

Abstract

BACKGROUND: Although optical microscopy (OM) remains the reference technique for analysis of ascitic (AF) and pleural (PF) fluids, novel hematological analyzers are equipped with modules for body fluid (BF) analysis. This study was aimed to analyze the performance of XN-BF module in Sysmex XN-9000, and to develop validation rules for automated cell counts in BFs.
METHODS: The evaluation of XN-BF module included assessment of carryover, Limit of Blank (LoB), Limit of Detection (LoD), Limit of Quantitation (LoQ), linearity, data comparison with OM, and development of rules for assisting the validation of automated analysis of BFs and activating reflex testing.
RESULTS: The carryover was negligible. The LoB, LoD, LoQ and linearity were always excellent. The comparison with OM was characterized by Pearson's correlations ranging from r=0.50 to r=0.99 (p<0.001), modest bias and high diagnostic concordance (Area Under the Curve between 0.85 and 0.99). The use of instrument-specific cut-offs further increased diagnostic concordance. The implementation of reflex testing rules based on XN-BF data increased sensitivity and specificity of BFs classification to 0.98 and 0.95.
CONCLUSIONS: Our results suggest that the XN-BF module on Sysmex-9000 may be a suitable alternative to OM for screening BF samples, especially when specific validation rules are used.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Ascitic fluid; Automated cell count; Body fluid; Pleural fluid; Reflex testing rules; Sysmex XN

Mesh:

Year:  2015        PMID: 26554518     DOI: 10.1016/j.cca.2015.11.005

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  4 in total

1.  Evaluation of Sysmex XN-1000 hematology analyzer for cell count and screening of malignant cells of serous cavity effusion.

Authors:  Weiyi Xu; Qian Yu; Lixia Xie; Baode Chen; Ling Zhang
Journal:  Medicine (Baltimore)       Date:  2017-07       Impact factor: 1.889

2.  Diagnostic Value of Pleural Effusion Mononuclear Cells Count and Adenosine Deaminase for Tuberculous Pleurisy Patients in China: A Case-Control Study.

Authors:  Xiaoli Lei; Junli Wang; Zhigang Yang; Shengli Zhou; Zhiwei Xu
Journal:  Front Med (Lausanne)       Date:  2019-12-17

3.  A comparison of the analysis of 3 types of body fluids using the XN-350 hematology analyzer versus light microscopy assessment.

Authors:  Jiwon Lee; Younggeun Cho; Han-Sung Kim; Hee Jung Kang; Miyoung Kim; Young Kyung Lee
Journal:  Medicine (Baltimore)       Date:  2021-03-19       Impact factor: 1.817

4.  Diagnostic Accuracy of Pleural Effusion Mononuclear Cells/Leukocyte Ratio in Tuberculous Pleurisy.

Authors:  Xiaoli Lei; Junli Wang; Zhigang Yang
Journal:  Front Med (Lausanne)       Date:  2021-03-18
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.