The origin of Korotkoff sounds and the accuracy of auscultatory blood pressure measurements.

Charles F Babbs1.   

Abstract

This study explores the hypothesis that the sharper, high frequency Korotkoff sounds come from resonant motion of the arterial wall, which begins after the artery transitions from a buckled state to an expanding state. The motions of one mass, two nonlinear springs, and one damper, driven by transmural pressure under the cuff, are used to model and compute the Korotkoff sounds according to principles of classical Newtonian physics. The natural resonance of this spring-mass-damper system provides a concise, yet rigorous, explanation for the origin of Korotkoff sounds. Fundamentally, wall stretching in expansion requires more force than wall bending in buckling. At cuff pressures between systolic and diastolic arterial pressure, audible vibrations (> 40 Hz) occur during early expansion of the artery wall beyond its zero pressure radius after the outward moving mass of tissue experiences sudden deceleration, caused by the discontinuity in stiffness between bucked and expanded states. The idealized spring-mass-damper model faithfully reproduces the time-domain waveforms of actual Korotkoff sounds in humans. Appearance of arterial sounds occurs at or just above the level of systolic pressure. Disappearance of arterial sounds occurs at or just above the level of diastolic pressure. Muffling of the sounds is explained by increased resistance of the artery to collapse, caused by downstream venous engorgement. A simple analytical model can define the physical origin of Korotkoff sounds, suggesting improved mechanical or electronic filters for their selective detection and confirming the disappearance of the Korotkoff sounds as the optimal diastolic end point.
Copyright © 2015 American Society of Hypertension. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Biomechanics; diastolic dilemma; muffling; sphygmomanometry

Mesh:

Year:  2015        PMID: 26553392     DOI: 10.1016/j.jash.2015.09.011

Source DB:  PubMed          Journal:  J Am Soc Hypertens        ISSN: 1878-7436


  8 in total

Review 1.  Arterial Stiffness in CKD: A Review.

Authors:  Raymond R Townsend
Journal:  Am J Kidney Dis       Date:  2018-06-19       Impact factor: 8.860

2.  Effects of Behaviorally Supported Exercise and Exercise-Induced Mood Changes on Elevated Blood Pressure and Hypertension in African American Adults with Severe Obesity.

Authors:  James J Annesi
Journal:  J Immigr Minor Health       Date:  2021-05-25

3.  Bio-Impedance Sensor for Real-Time Artery Diameter Waveform Assessment.

Authors:  Mugeb Al-Harosh; Marat Yangirov; Dmitry Kolesnikov; Sergey Shchukin
Journal:  Sensors (Basel)       Date:  2021-12-17       Impact factor: 3.576

Review 4.  Korotkoff sounds dynamically reflect changes in cardiac function based on deep learning methods.

Authors:  Wenting Lin; Sixiang Jia; Yiwen Chen; Hanning Shi; Jianqiang Zhao; Zhe Li; Yiteng Wu; Hangpan Jiang; Qi Zhang; Wei Wang; Yayu Chen; Chao Feng; Shudong Xia
Journal:  Front Cardiovasc Med       Date:  2022-08-26

5.  Changes in Office Blood Pressure Control, Augmentation Index, and Liver Steatosis in Kidney Transplant Patients after Successful Hepatitis C Infection Treatment with Direct Antiviral Agents.

Authors:  Aureliusz Kolonko; Joanna Musialik; Jerzy Chudek; Magdalena Bartmańska; Natalia Słabiak-Błaż; Agata Kujawa-Szewieczek; Piotr Kuczera; Katarzyna Kwiecień-Furmańczuk; Andrzej Więcek
Journal:  J Clin Med       Date:  2020-03-30       Impact factor: 4.241

6.  Novel methods of testing and calibration of oscillometric blood pressure monitors.

Authors:  Branko G Celler; Ahmadreza Argha; Phu Ngoc Le; Eliathamby Ambikairajah
Journal:  PLoS One       Date:  2018-08-06       Impact factor: 3.240

7.  Development of a Multi-Array Pressure Sensor Module for Radial Artery Pulse Wave Measurement.

Authors:  Donggeun Roh; Sangjin Han; Junyung Park; Hangsik Shin
Journal:  Sensors (Basel)       Date:  2019-12-19       Impact factor: 3.576

8.  Design and Fabrication of a New Wearable Pressure Sensor for Blood Pressure Monitoring.

Authors:  Marian Ion; Silviu Dinulescu; Bogdan Firtat; Mihaela Savin; Octavian N Ionescu; Carmen Moldovan
Journal:  Sensors (Basel)       Date:  2021-03-16       Impact factor: 3.576

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.