From bench to bedside: Can the improvements in left ventricular assist device design mitigate adverse events and increase survival?

Vincenzo Tarzia1, Gabriele Di Giammarco2, Michele Di Mauro2, Giacomo Bortolussi3, Massimo Maccherini4, Vincenzo Tursi5, Massimo Maiani5, Sonia Bernazzali4, Daniele Marinelli2, Massimiliano Foschi2, Edward Buratto3, Jonida Bejko3, Dario Gregori3, Silvia Scuri3, Ugolino Livi5, Guido Sani4, Tomaso Bottio3, Gino Gerosa3.   

Abstract

OBJECTIVE: In vitro tests demonstrated that the new cone-bearing configuration of the Jarvik 2000 (Jarvik Heart Inc, New York, NY) left ventricular assist device exhibits better hydraulic efficiency than the previous pin-bearing design. We investigated the long-term outcomes of patients who received the Jarvik 2000 left ventricular assist device, depending on bearing design.
METHODS: A retrospective review of prospectively collected data from 18 centers included in the Italian Registry was performed. From May 2008 to September 2013, 99 patients with end-stage heart failure were enrolled. Patients were divided into 2 groups according to their Jarvik 2000 suspending mechanism: Group pin included patients with pin bearings (May 2008 to June 2010), and group cone included patients with newer cone bearings (July 2010 to September 2013). The 2 groups did not differ significantly in terms of baseline characteristics.
RESULTS: A total of 30 of 39 patients (group pin) and 46 of 60 patients (group cone) were discharged. During follow-up, 6 patients underwent transplantation, and in 1 patient the left ventricular assist device was explanted. The cumulative incidence competing risk of the entire cohort for noncardiovascular-related death was 28% (20%-40%); the cumulative incidence competing risk for cardiovascular-related death was 56% (42%-73%): 71% in group pin versus 26% in group cone (P = .034). The multivariate analyses confirmed that the pin-bearing design was a risk factor for cardiovascular death, along with Interagency Registry for Mechanically Assisted Circulatory Support class. Right ventricular failures and ischemic and hemorrhagic strokes were significantly higher in group pin.
CONCLUSIONS: Patients with the new pump configuration showed a better freedom from cardiovascular death and lower incidence of fatal stroke and right ventricular failure. Further studies are needed to prove the favorable impact of pump-enhanced fluid dynamics on long-term results.
Copyright © 2016 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Jarvik 2000; LVAD; bearing system; follow-up; heart failure; ventricular assist device

Mesh:

Year:  2015        PMID: 26548997     DOI: 10.1016/j.jtcvs.2015.09.107

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  1 in total

1.  Sternal-sparing Jarvik 2000 implantation, heart transplantation and driveline pedestal removal.

Authors:  Antonio Loforte; Gregorio Gliozzi; Giuditta Coppola; Giulio Giovanni Cavalli; Sofia Martin Suarez; Davide Pacini
Journal:  Ann Cardiothorac Surg       Date:  2021-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.