On the design of a DEA-based device to pot entially assist lower leg disorders: an analytical and FEM investigation accounting for nonlinearities of the leg and device deformations.

Shahram Pourazadi1, Sadegh Ahmadi2, Carlo Menon3.   

Abstract

BACKGROUND: One of the recommended treatments for disorders associated with the lower extremity venous insufficiency is the application of external mechanical compression. Compression stockings and elastic bandages are widely used for the purpose of compression therapy and are usually designed to exert a specified value or range of compression on the leg. However, the leg deforms under external compression, which can lead to undesirable variations in the amount of compression applied by the compression bandages. In this paper, the use of an active compression bandage (ACB), whose compression can be regulated through an electrical signal, is investigated. The ACB is based on the use of dielectric elastomer actuators. This paper specifically investigates, via both analytical and non-linear numerical simulations, the potential pressure the ACB can apply when the compliancy of the human leg is taken into account. The work underpins the need to account for the compressibility of the leg when designing compression garments for lower extremity venous insufficiency.
METHODS: A mathematical model is used to simulate the volumetric change of a calf when compressed. Suitable parameters for this calf model are selected from the literature where the calf, from ankle to knee, is divided into six different regions. An analytical electromechanical model of the ACB, which considers its compliancy as a function of its pre-stretch and electricity applied, is used to predict the ACB's behavior. Based on these calf and ACB analytical models, a simulation is performed to investigate the interaction between the ACB and the human calf with and without an electrical stimulus applied to the ACB. This simulation is validated by non-linear analysis performed using a software based on the finite element method (FEM). In all simulations, the ACB's elastomer is stretched to a value in the range between 140 and 220 % of its initial length.
RESULTS: Using data from the literature, the human calf model, which is examined in this work, has different compliancy in its different regions. For example, when a 28.5 mmHg (3.8 kPa) of external compression is applied to the entire calf, the ankle shows a 3.7 % of volume change whereas the knee region undergoes a 2.7 % of volume change. The paper presents the actual pressure in the different regions of the calf for different values of the ACB's stretch ratio when it is either electrically activated or not activated, and when compliancy of the leg is either considered or not considered. For example, results of the performed simulation show that about 10 % variation in compression in the ankle region is expected when the ACB initially applies 6 kPa and the compressibility of the calf is first considered and then not considered. Such a variation reduces to 5 % when the initial pressure applied by the ACB reduced by half.
CONCLUSIONS: Comparison with non-linear FEM simulations show that the analytical models used in this work can closely estimate interaction between an active compression bandage and a human calf. In addition, compliancy of the leg should not be neglected when either designing a compression band or predicting the compressive force it can exert. The methodology proposed in this work can be extended to other types of elastic compression bandages and garments for biomedical applications.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26541150      PMCID: PMC4635664          DOI: 10.1186/s12938-015-0088-3

Source DB:  PubMed          Journal:  Biomed Eng Online        ISSN: 1475-925X            Impact factor:   2.819


  19 in total

1.  Prevention of deep-vein thrombosis after total hip and knee replacement. Low-molecular-weight heparin in combination with intermittent pneumatic compression.

Authors:  Y Silbersack; B M Taute; W Hein; H Podhaisky
Journal:  J Bone Joint Surg Br       Date:  2004-08

2.  Type of compression for reducing venous stasis. A study of lower extremities during inactive recumbency.

Authors:  B Sigel; A L Edelstein; L Savitch; J H Hasty; W R Felix
Journal:  Arch Surg       Date:  1975-02

3.  Changes in venous blood volume produced by external compression of the lower leg.

Authors:  R B Thirsk; R D Kamm; A H Shapiro
Journal:  Med Biol Eng Comput       Date:  1980-09       Impact factor: 2.602

4.  Efficacy of graded elastic compression in the lower leg.

Authors:  S Pierson; D Pierson; R Swallow; G Johnson
Journal:  JAMA       Date:  1983-01-14       Impact factor: 56.272

5.  A physiological study of elastic compression stockings in venous disorders of the leg.

Authors:  N A Jones; P J Webb; R I Rees; V V Kakkar
Journal:  Br J Surg       Date:  1980-08       Impact factor: 6.939

6.  Effect of positive external pressure on calf volume and local venous haemodynamics.

Authors:  M Zicot; K H Parker; C G Caro
Journal:  Phys Med Biol       Date:  1977-11       Impact factor: 3.609

7.  Prevention and treatment of postphlebitic syndrome: results of a 3-part study.

Authors:  J S Ginsberg; J Hirsh; J Julian; M Vander LaandeVries; D Magier; B MacKinnon; M Gent
Journal:  Arch Intern Med       Date:  2001-09-24

8.  Smaller age-associated reductions in leg venous compliance in endurance exercise-trained men.

Authors:  K D Monahan; F A Dinenno; D R Seals; J R Halliwill
Journal:  Am J Physiol Heart Circ Physiol       Date:  2001-09       Impact factor: 4.733

9.  Measurement of limb venous compliance in humans: technical considerations and physiological findings.

Authors:  J R Halliwill; C T Minson; M J Joyner
Journal:  J Appl Physiol (1985)       Date:  1999-10

10.  Are compression stockings an effective treatment for orthostatic presyncope?

Authors:  Clare Louise Protheroe; Anastasia Dikareva; Carlo Menon; Victoria Elizabeth Claydon
Journal:  PLoS One       Date:  2011-12-16       Impact factor: 3.240

View more
  1 in total

1.  Intermittent Calf Compression Delays the Onset of Presyncope in Young Healthy Individuals.

Authors:  Brooke C D Hockin; Victoria E Claydon
Journal:  Front Physiol       Date:  2020-01-23       Impact factor: 4.566

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.