Numerical and clinical precision of continuous glucose monitoring in Colombian patients treated with insulin infusion pump with automated suspension in hypoglycemia.

Ana M Gómez1, Alejandro Marín Sánchez1, Oscar M Muñoz2, Christian Alejandro Colón Peña1.   

Abstract

OBJECTIVE: Insulin pump therapy associated with continuous glucose monitoring has shown a positive clinical impact on diabetes control and reduction of hypoglycemia episodes. There are descriptions of the performance of this device in other populations, but its precision and accuracy in Colombia and Latin America are unknown, especially in the routine outpatient setting.
METHODS: Data from 33 type 1 and type 2 diabetes patients with sensor-augmented pump therapy with threshold suspend automation, MiniMed Paradigm® Veo™ (Medtronic, Northridge, California), managed at Hospital Universitario San Ignacio (Bogotá, Colombia) and receiving outpatient treatment, were analyzed. Simultaneous data from continuous glucose monitoring and capillary blood glucose were compared, and their precision and accuracy were calculating with different methods, including Clarke error grid.
RESULTS: Analyses included 2,262 continuous glucose monitoring -reference paired glucose values. A mean absolute relative difference of 20.1% was found for all measurements, with a value higher than 23% for glucose levels ≤75mg/dL. Global compliance with the ISO criteria was 64.9%. It was higher for values >75mg/dl (68.3%, 1,308 of 1,916 readings), than for those ≤ 75mg/dl (49.4%, 171 of 346 readings). Clinical accuracy, as assessed by the Clarke error grid, showed that 91.77% of data were within the A and B zones (75.6% in hypoglycemia).
CONCLUSIONS: A good numerical accuracy was found for continuous glucose monitoring in normo and hyperglycemia situations, with low precision in hypoglycemia. The clinical accuracy of the device was adequate, with no significant safety concerns for patients.
Copyright © 2015 SEEN. Published by Elsevier España, S.L.U. All rights reserved.

Entities:  

Keywords:  Bomba de insulina; Continuous glucose monitoring; Diabetes tipo 1; Diabetes tipo 2; Insulin pump; Monitorización continua de glucosa; Type 1 diabetes mellitus; Type 2 diabetes mellitus

Mesh:

Substances:

Year:  2015        PMID: 26531841     DOI: 10.1016/j.endonu.2015.06.012

Source DB:  PubMed          Journal:  Endocrinol Nutr        ISSN: 1575-0922


  2 in total

Review 1.  Devices for continuous monitoring of glucose: update in technology.

Authors:  Ana María Gómez; Diana Cristina Henao Carrillo; Oscar Mauricio Muñoz Velandia
Journal:  Med Devices (Auckl)       Date:  2017-09-12

2.  Survey on hypoglycemia among insulin-treated patients with diabetes: The Colombian International Operations Hypoglycemia Assessment Tool population

Authors:  Ana María Gómez; Luis G Chica; Álvaro F Burbano; Esdras M Vásquez; Jorge A Escobar; Paola M Arias; Dora I Molina
Journal:  Biomedica       Date:  2019-09-01       Impact factor: 0.935

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.