The edge artifact in the point-spread function-based PET reconstruction at different sphere-to-background ratios of radioactivity.

Daisuke Kidera1, Ken Kihara1, Go Akamatsu1, Shohei Mikasa1, Takafumi Taniguchi1, Yuji Tsutsui2, Toshiki Takeshita1, Akira Maebatake1, Kenta Miwa1, Masayuki Sasaki3.   

Abstract

OBJECTIVE: The aim of this study was to quantitatively evaluate the edge artifacts in PET images reconstructed using the point-spread function (PSF) algorithm at different sphere-to-background ratios of radioactivity (SBRs).
METHODS: We used a NEMA IEC body phantom consisting of six spheres with 37, 28, 22, 17, 13 and 10 mm in inner diameter. The background was filled with (18)F solution with a radioactivity concentration of 2.65 kBq/mL. We prepared three sets of phantoms with SBRs of 16, 8, 4 and 2. The PET data were acquired for 20 min using a Biograph mCT scanner. The images were reconstructed with the baseline ordered subsets expectation maximization (OSEM) algorithm, and with the OSEM + PSF correction model (PSF). For the image reconstruction, the number of iterations ranged from one to 10. The phantom PET image analyses were performed by a visual assessment of the PET images and profiles, a contrast recovery coefficient (CRC), which is the ratio of SBR in the images to the true SBR, and the percent change in the maximum count between the OSEM and PSF images (Δ % counts).
RESULTS: In the PSF images, the spheres with a diameter of 17 mm or larger were surrounded by a dense edge in comparison with the OSEM images. In the spheres with a diameter of 22 mm or smaller, an overshoot appeared in the center of the spheres as a sharp peak in the PSF images in low SBR. These edge artifacts were clearly observed in relation to the increase of the SBR. The overestimation of the CRC was observed in 13 mm spheres in the PSF images. In the spheres with a diameter of 17 mm or smaller, the Δ % counts increased with an increasing SBR. The Δ % counts increased to 91 % in the 10-mm sphere at the SBR of 16.
CONCLUSIONS: The edge artifacts in the PET images reconstructed using the PSF algorithm increased with an increasing SBR. In the small spheres, the edge artifact was observed as a sharp peak at the center of spheres and could result in overestimation.

Keywords:  Edge artifact; PET; Point-spread function

Mesh:

Year:  2015        PMID: 26531181     DOI: 10.1007/s12149-015-1036-9

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  4 in total

1.  Edge Artifacts in Point Spread Function-based PET Reconstruction in Relation to Object Size and Reconstruction Parameters.

Authors:  Yuji Tsutsui; Shinichi Awamoto; Kazuhiko Himuro; Yoshiyuki Umezu; Shingo Baba; Masayuki Sasaki
Journal:  Asia Ocean J Nucl Med Biol       Date:  2017

2.  New standards for phantom image quality and SUV harmonization range for multicenter oncology PET studies.

Authors:  Go Akamatsu; Naoki Shimada; Keiichi Matsumoto; Hiromitsu Daisaki; Kazufumi Suzuki; Hiroshi Watabe; Keiichi Oda; Michio Senda; Takashi Terauchi; Ukihide Tateishi
Journal:  Ann Nucl Med       Date:  2022-01-14       Impact factor: 2.668

3.  Evaluation of spatial dependence of point spread function-based PET reconstruction using a traceable point-like 22Na source.

Authors:  Taisuke Murata; Kenta Miwa; Noriaki Miyaji; Kei Wagatsuma; Tomoyuki Hasegawa; Keiichi Oda; Takuro Umeda; Takashi Iimori; Yoshitada Masuda; Takashi Terauchi; Mitsuru Koizumi
Journal:  EJNMMI Phys       Date:  2016-10-26

4.  Multicentre analysis of PET SUV using vendor-neutral software: the Japanese Harmonization Technology (J-Hart) study.

Authors:  Yuji Tsutsui; Hiromitsu Daisaki; Go Akamatsu; Takuro Umeda; Matsuyoshi Ogawa; Hironori Kajiwara; Shigeto Kawase; Minoru Sakurai; Hiroyuki Nishida; Keiichi Magota; Kazuaki Mori; Masayuki Sasaki
Journal:  EJNMMI Res       Date:  2018-08-20       Impact factor: 3.138

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.