Changing Default Fluoroscopy Equipment Settings Decreases Entrance Skin Dose in Patients.

Benjamin K Canales1, Lindsay Sinclair2, Diana Kang3, Anna M Mench4, Manuel Arreola5, Vincent G Bird6.   

Abstract

PURPOSE: Proper fluoroscopic education and protocols may reduce the patient radiation dose but few prospective studies in urology have been performed. Using optically stimulated luminescent dosimeters we tested whether fluoroscopy time and/or entrance skin dose would decrease after educational and radiation reduction protocols.
MATERIALS AND METHODS: At default manufacturer settings fluoroscopy time and entrance skin dose were prospectively measured using optically stimulated luminescent dosimeters in patients undergoing ureteroscopy, retrograde pyelogram/stent or percutaneous nephrolithotomy with access for stone disease. A validated radiation safety competency test was administered to urology faculty and residents before and after web based, hands-on fluoroscopy training. Default fluoroscopy settings were changed from continuous to intermittent pulse rate and from standard to half-dose output. Fluoroscopy time and entrance skin dose were then measured again.
RESULTS: The cohorts of 44 pre-protocol and 50 post-protocol patients with stones were similarly matched. The change in mean fluoroscopy time and entrance skin dose from pre-protocol to post-protocol was -0.6 minutes and -11.6 mGy (33%) for percutaneous nephrolithotomy (p = 0.62 and <0.001), 0.5 minutes and -0.1 mGy (34%) for ureteroscopy (p = 0.42 and 0.31), and 0.1 minute and -0.1 mGy (29%) for retrograde pyelogram/stent (p = 0.85 and 0.49, respectively). Urologist post-training test scores increased 30% from pretraining scores (p = 0.1).
CONCLUSIONS: Radiation safety training protocols improved clinical knowledge but did not significantly alter fluoroscopy time. Changing equipment default settings to intermittent pulse rate (12 frames per second) and half-dose lowered the entrance skin dose by 30% across all endourology patients but most significantly during percutaneous nephrolithotomy. To limit patient radiation exposure fluoroscopy default settings should be decreased before all endourology procedures and image equipment manufacturers should consider lowering standard default renal settings.
Copyright © 2016 American Urological Association Education and Research, Inc. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  fluoroscopy; kidney; nephrolithiasis; radiation dosage; skin

Mesh:

Year:  2015        PMID: 26527512     DOI: 10.1016/j.juro.2015.10.088

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450