Stabilization of Coronary Stents Using the Floating-Wire Technique.

Yue Li, Jingyi Xue, Shuang Li, Li Sheng, Danghui Sun, Weimin Li, Yongtai Gong1.   

Abstract

AIMS: Excessive movement of coronary stents within the artery can make accurate stent placement difficult. This study assessed the use of the floating-wire technique to reduce stent motion to improve placement accuracy. METHODS AND
RESULTS: During percutaneous coronary intervention, if excessive stent movement prevented accurate stent placement, the floating-wire technique was performed to reduce stent motion within the coronary artery during both stent positioning and deployment. Postprocedural angiograms were analyzed by two independent operators to measure the motion length of the stent delivery system relative to the coronary artery before and after using the floating-wire technique. The floating-wire technique was considered necessary in 19 patients. No procedural complications occurred. The extent of motion was markedly reduced by using the floating-wire technique (4.04 ± 1.25 mm to 1.11 ± 0.81 mm; P<.001).
CONCLUSIONS: The floating-wire technique is an effective and safe method to reduce stent movement and facilitate accurate stent deployment. This simple technique can be easily applied in any interventional cardiac catheterization laboratory without the need for additional training or equipment.

Entities:  

Mesh:

Year:  2015        PMID: 26524202

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  1 in total

1.  The Use of Bumper Wire Technique and Intravascular Ultrasound for Precise Aorto-Ostial Stenting.

Authors:  Pavan K V Reddy; Joseph Daibes; Michel Skaf; Roberto Ochoa; Tomohiro Fujisaki; Patricia Lin; Apurva Patel; Tak Kwan
Journal:  Front Cardiovasc Med       Date:  2022-07-14