Technical Note: Immunohistochemical evaluation of mouse brain irradiation targeting accuracy with 3D-printed immobilization device.

Niloufar Zarghami1, Michael D Jensen1, Srikanth Talluri2, Paula J Foster3, Ann F Chambers4, Frederick A Dick2, Eugene Wong5.   

Abstract

PURPOSE: Small animal immobilization devices facilitate positioning of animals for reproducible imaging and accurate focal radiation therapy. In this study, the authors demonstrate the use of three-dimensional (3D) printing technology to fabricate a custom-designed mouse head restraint. The authors evaluate the accuracy of this device for the purpose of mouse brain irradiation.
METHODS: A mouse head holder was designed for a microCT couch using cad software and printed in an acrylic based material. Ten mice received half-brain radiation while positioned in the 3D-printed head holder. Animal placement was achieved using on-board image guidance and computerized asymmetric collimators. To evaluate the precision of beam localization for half-brain irradiation, mice were sacrificed approximately 30 min after treatment and brain sections were stained for γ-H2AX, a marker for DNA breaks. The distance and angle of the γ-H2AX radiation beam border to longitudinal fissure were measured on histological samples. Animals were monitored for any possible trauma from the device.
RESULTS: Visualization of the radiation beam on ex vivo brain sections with γ-H2AX immunohistochemical staining showed a sharp radiation field within the tissue. Measurements showed a mean irradiation targeting error of 0.14±0.09 mm (standard deviation). Rotation between the beam axis and mouse head was 1.2°±1.0° (standard deviation). The immobilization device was easily adjusted to accommodate different sizes of mice. No signs of trauma to the mice were observed from the use of tooth block and ear bars.
CONCLUSIONS: The authors designed and built a novel 3D-printed mouse head holder with many desired features for accurate and reproducible radiation targeting. The 3D printing technology was found to be practical and economical for producing a small animal imaging and radiation restraint device and allows for customization for study specific needs.

Entities:  

Mesh:

Year:  2015        PMID: 26520740     DOI: 10.1118/1.4933200

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Half brain irradiation in a murine model of breast cancer brain metastasis: magnetic resonance imaging and histological assessments of dose-response.

Authors:  Niloufar Zarghami; Donna H Murrell; Michael D Jensen; Frederick A Dick; Ann F Chambers; Paula J Foster; Eugene Wong
Journal:  Radiat Oncol       Date:  2018-06-01       Impact factor: 3.481

2.  A review of 3D printed patient specific immobilisation devices in radiotherapy.

Authors:  Amirhossein Asfia; James I Novak; Mazher Iqbal Mohammed; Bernard Rolfe; Tomas Kron
Journal:  Phys Imaging Radiat Oncol       Date:  2020-03-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.